Longlands M G, Wiener K
Ann Clin Biochem. 1978 May;15(3):164-7. doi: 10.1177/000456327801500136.
Differences in alkaline phosphatase activity of some commerical control sera were observed when comparing manual with AutoAnalyzer versions of the Kind and King procedure. This phenomenon was found to be associated with sera boosted with exogenous alkaline phosphatase and to be owing mainly to inadequate provision of buffer in some automated methods. It is suggested that a modification to the SMA procedure may lessen the problem, but the constraints of the system necessitate a compromise. There was better precision after the modification had been introduced. Awareness of this phenomenon is of paramount importance if commerical sera are to be used in any way in the calibration of automated systems for alkaline phosphatase assay.
在比较金氏法手工操作版本和自动分析仪版本时,观察到一些商业对照血清的碱性磷酸酶活性存在差异。发现这种现象与用外源性碱性磷酸酶增强的血清有关,并且主要是由于某些自动化方法中缓冲液供应不足。建议对SMA法进行修改可能会减轻该问题,但系统的限制需要进行折衷。引入修改后精度有所提高。如果要在碱性磷酸酶测定的自动化系统校准中以任何方式使用商业血清,认识到这种现象至关重要。