Suppr超能文献

使用光离子化检测器分析用于检测二硫化碳的活性炭管样品。

Analysis of charcoal tube samples for carbon disulfide using a photoionization detector.

作者信息

Smith D B, Krause L A

出版信息

Am Ind Hyg Assoc J. 1978 Dec;39(12):939-44. doi: 10.1080/0002889778507892.

Abstract

A gas chromatograph equipped with a photoionization detector was used to detect carbon disulfide eluted from charcoal tubes with acetonitrile. Overall sampling and desorption efficiency for carbon disulfide was found to approach 100 percent when compared to a liquid impingement, colorimetric method. A relative standard deviation of 4.78 percent (sampling plus analysis) was observed when the method was applied to a synthetic test atmosphere of 10 parts per million carbon disulfide. Sensitivity of the proposed method will enable concentrations of 0.10 ppm CS2 to be determined in a 10 liter air sample. The photoionization detector offers several advantages over other detectors, namely ease of operation and greater sensitivity.

摘要

使用配备光离子化检测器的气相色谱仪检测用乙腈从活性炭管中洗脱出来的二硫化碳。与液体冲击比色法相比,二硫化碳的整体采样和解吸效率接近100%。当该方法应用于百万分之十的二硫化碳合成测试气氛时,观察到相对标准偏差为4.78%(采样加分析)。所提出方法的灵敏度能够在10升空气样品中测定0.10 ppm的二硫化碳浓度。与其他检测器相比,光离子化检测器具有几个优点,即操作简便且灵敏度更高。