Curtis G P, Lief L H
J Electrocardiol. 1980 Oct;13(4):401-4. doi: 10.1016/s0022-0736(80)80094-x.
P wave triggered pacemakers can produce complicated arrhythmias during normal function. A ventricular, R wave inhibited pacemaker and an atrial sensing pacemaker were both present and functioning normally in the patient reported here. A premature paced beat was observed every 12th beat which resulted from discharge of the atrial sensing pacemaker. This arrhythmia appeared because a "P" wave occurred during the noise sampling period of the atrial sensing pacemaker, disabling the demand function and resulting in generator discharge on or near the T wave of a preceding unsensed beat while the pacemaker was in the asynchronous mode.