Pöge A W, Otto U
Z Gesamte Inn Med. 1980 May 1;35(9):375-8.
The changes of the acid-base- and blood gas parameters were continuously measured under different storage conditions by means of the blood-gas automaton ABL 1 with the help of 40 blood tests of various patients. The results are discussed and a recommendation for practice is given. Apart from the necessary immediate determinations (without any restriction) in the hermetically closed blood tests transported or stored at 0--5 degrees C even three hours after blood taking analyses were got which could still be used clinically. In these cases by the clinicians only a decrease of BE of 2 mmol/l of blood should be taken into consideration. The result render possible a greater usability of the imported blood gas automatons.
借助ABL 1血气自动分析仪,通过对不同患者进行40次血液检测,持续测量了在不同储存条件下酸碱和血气参数的变化。对结果进行了讨论并给出了实际应用建议。除了在0-5摄氏度运输或储存的密闭血液检测中进行必要的即时测定(无任何限制)外,即使在采血后三小时进行分析,仍可获得可用于临床的结果。在这些情况下,临床医生只需考虑血液碱剩余(BE)降低2 mmol/l。这些结果使得进口血气自动分析仪具有更高的可用性。