The effect of mAs variation upon computed tomography image quality as evaluated by in vivo and in vitro studies.

作者信息

Haaga J R, Miraldi F, MacIntyre W, LiPuma J P, Bryan P J, Wiesen E

出版信息

Radiology. 1981 Feb;138(2):449-54. doi: 10.1148/radiology.138.2.7455129.

Abstract

The effect of different mAs settings on computed tomobraphy (CT) image quality was evaluated by a clinical study and phantom studies. In the clinical portion of this study, a number of CT images, obtained at various mAs settings, were reviewed by experienced CT interpreters. From these same images, graphs were charted of the root-mean-standard deviation (RMSD) vs. body size; there appeared to be a linear relationship between noise, as measured by RMSD, and body size. A performance phantom was used to evaluate the effect of the different mAs settings on resolution. There was some loss in low contrast resolution at 6 and 20 mAs; however, the resolution was identical at 40 and 100 mAs.

摘要

通过一项临床研究和体模研究评估了不同毫安秒(mAs)设置对计算机断层扫描(CT)图像质量的影响。在本研究的临床部分,由经验丰富的CT解读人员对在各种mAs设置下获得的一些CT图像进行了评估。从这些相同的图像中,绘制了均方根标准差(RMSD)与身体大小的关系图;结果显示,以RMSD衡量的噪声与身体大小之间存在线性关系。使用性能体模来评估不同mAs设置对分辨率的影响。在6 mAs和20 mAs时,低对比度分辨率有一定损失;然而,在40 mAs和100 mAs时,分辨率是相同的。