Suppr超能文献

作为阈值动态测量一部分的自动音调衰减测试。

The automatic tone decay test as part of dynamic measurement of threshold.

作者信息

Lehnhardt E, Battmer R D

出版信息

Arch Otorhinolaryngol. 1980;226(3):155-60. doi: 10.1007/BF00455130.

Abstract

Only dynamic recording of the tone threshold can give information on pathologic adaptation, auditory fatigue, and aggravation. We have therefore automated the tone decay test. This was achieved by modifying the Grason-Stadler audiometer E 800 and, concurrently, by developing a special device with integrated circuits for this purpose. The print-out is made by a commerical x-y-recorder which is also used for recording the impedance test results. During the automatic tone decay test the patient himself increases the intensity of the tone only until he can hear the tone again, the intensity of the tone otherwise remaining constant. Thus, we obtain either a flat (adaptation) or a sharply declining (auditory fatigue) curve which correlates with the Békésy continuous tone curve, but may show a distinctive separation. The tone decay test is - in contrast to all other suprathreshold tests - independent of the amount of hearing loss. It can therefore give information on the kind of hearing impairment for normal short duration thresholds as well.

摘要

只有动态记录音调阈值才能提供有关病理适应、听觉疲劳和病情加重的信息。因此,我们已将音调衰减测试自动化。这是通过改装Grason-Stadler E 800听力计,并同时为此开发一种带有集成电路的特殊装置来实现的。打印输出由一台商用x-y记录仪完成,该记录仪也用于记录阻抗测试结果。在自动音调衰减测试期间,患者自己仅将音调强度增加到再次能听到该音调时为止,否则音调强度保持恒定。这样,我们就得到一条平坦的(适应)或急剧下降的(听觉疲劳)曲线,它与贝凯西连续音调曲线相关,但可能会显示出明显的分离。与所有其他阈上测试不同,音调衰减测试与听力损失的程度无关。因此,它也可以为正常短持续时间阈值的听力损伤类型提供信息。

文献检索

告别复杂PubMed语法,用中文像聊天一样搜索,搜遍4000万医学文献。AI智能推荐,让科研检索更轻松。

立即免费搜索

文件翻译

保留排版,准确专业,支持PDF/Word/PPT等文件格式,支持 12+语言互译。

免费翻译文档

深度研究

AI帮你快速写综述,25分钟生成高质量综述,智能提取关键信息,辅助科研写作。

立即免费体验