van Leersum M D, Schweitzer M E, Gannon F, Vinitski S, Finkel G, Mitchell D G
Department of Radiology, Thomas Jefferson University Hospital, Philadelphia, Pennsylvania, USA.
Skeletal Radiol. 1995 Aug;24(6):431-5. doi: 10.1007/BF00941240.
Since the thickness of cartilage is an important indicator of the status, progression and response to therapy of articular disorders, assessment of it is desirable. This study was undertaken to assess the accuracy, precision, and reliability of magnetic resonance (MR) measurements of articular cartilage.
Fifteen cadaveric patellas were imaged in the axial plane at 1.5 T. Gradient echo and fat-suppressed FSE, T2-weighted, proton density, and T1-weighted sequences were performed. We measured each 5-mm section separately at three standardized positions, giving a total of 900 measurements. These findings were correlated with independently performed measurements of the corresponding anatomic sections. A hundred random measurements were also evaluated for reproducibility and interobserver variation.
Although all sequences were highly accurate (range r = 0.78-0.82), the T1-weighted images were the most accurate, with a mean difference of 0.25 mm and a correlation coefficient of 0.85. All sequences were also highly reproducible (mean difference between -0.09 and 0.05 mm) with little inter-observer variation (mean difference -0.04 and 0.11 mm). In an attempt to improve the accuracy of the MR measurements further, we retrospectively evaluated all measurements with discrepancies greater than 1 mm from the specimen. All these differences were attributable to focal defects causing exaggeration of the thickness on MR imaging.
MR imaging is accurate, precise, and reliable as a basis for measuring articular cartilage and may potentially be usable to monitor progression of articular disorders. Care must be taken not to overestimate the thickness of areas with surface defects.
由于软骨厚度是关节疾病状态、进展及治疗反应的重要指标,因此对其进行评估很有必要。本研究旨在评估磁共振(MR)测量关节软骨的准确性、精密度和可靠性。
对15个尸体髌骨在1.5T场强下进行轴向平面成像。采用梯度回波和脂肪抑制快速自旋回波、T2加权、质子密度加权及T1加权序列。我们在三个标准化位置分别测量每5毫米的切片,共进行900次测量。将这些结果与相应解剖切片的独立测量结果进行相关性分析。还对100次随机测量进行了重复性和观察者间差异评估。
尽管所有序列的准确性都很高(相关系数r范围为0.78 - 0.82),但T1加权图像最为准确,平均差异为0.25毫米,相关系数为0.85。所有序列的重复性也都很高(平均差异在 - 0.09至0.05毫米之间),观察者间差异很小(平均差异为 - 0.04至0.11毫米)。为进一步提高MR测量的准确性,我们回顾性评估了所有与标本差异大于1毫米的测量结果。所有这些差异均归因于导致MR成像上厚度夸大的局灶性缺损。
MR成像作为测量关节软骨的基础是准确、精密且可靠的,可能可用于监测关节疾病的进展。必须注意不要高估存在表面缺损区域的厚度。