[诊断性CT中的辐射暴露:螺旋CT与标准CT的比较]

[Radiation exposure in diagnostic CT: comparison between spiral CT and standard CT].

作者信息

Pitton M B, Harsini M, Mohr W, Schweden F, Düber C

机构信息

Klinik für Radiologie, Universitätsklinik, Mainz.

出版信息

Aktuelle Radiol. 1995 Sep;5(5):289-92.

PMID:7495889
Abstract

CT-Dose-Index of Standard-CT corresponds to CT-Dose-Effect-Index of Spiral-CT. Spiral-CT provides a continuous data acquisition with well balanced dose profiles. Increasing pitch reduces radiation dose and nevertheless an unbroken reconstruction of scanned volume is possible. In contrast to Standard-CT, small parts at both ends of the volume scan are not reconstructed. Volume-Dose-Index therefore is elevated in short scan times but corresponds to that of Standard-CT in long scan times. 1 mm slices cause more radiation exposure due to their broadened dose profiles.

摘要

标准CT的CT剂量指数与螺旋CT的CT剂量效应指数相对应。螺旋CT提供连续的数据采集,剂量分布平衡良好。增加螺距可降低辐射剂量,并且仍有可能对扫描容积进行不间断的重建。与标准CT不同,容积扫描两端的小部位不进行重建。因此,容积剂量指数在短扫描时间内会升高,但在长扫描时间内与标准CT的相当。1毫米层厚由于其加宽的剂量分布会导致更多的辐射暴露。