Giacomuzzi S M, Erckert B, Freund M C, Schöpf T, Dessl A, Jaschke W
Univ.-Klinik für Radiodiagnostik 1, Univ.-Klinik Innsbruck.
Aktuelle Radiol. 1996 Mar;6(2):110-3.
A new investigation technique in computed tomography--the Smart Scan process--and the associated reduction in tube current are discussed. In addition, the reduced dosage values resulting from the reduced current values are compared with those of a standard measurement and image characteristics such as picture unit noise are evaluated. By means of three special water phantoms with dimensions corresponding to those of actual patient geometries and a pool of 183 patients, the Smart Scan process was tested by the control algorithm implemented on a spiral computer tomograph. The dosage values with and without the Smart Scan being activated were determined. Dosage reductions of up to 18% can be realized with this new examination technique. In particular, patients with transverse oval head profiles will benefit from this modality.