Suppr超能文献

美国药典溶出度测定仪3(往复圆筒型):仪器参数对缓释制剂药物释放的影响

USP dissolution apparatus 3 (reciprocating cylinder): instrument parameter effects on drug release from sustained release formulations.

作者信息

Rohrs B R, Burch-Clark D L, Witt M J, Stelzer D J

机构信息

Analytical Research and Specifications Development, Upjohn Company, Kalamazoo, MI, USA.

出版信息

J Pharm Sci. 1995 Aug;84(8):922-6. doi: 10.1002/jps.2600840804.

Abstract

The United States Pharmacopoeia dissolution apparatus 3 (reciprocating cylinder) was evaluated with respect to effects of changes in instrument parameters on drug release rate from six hydrophilic matrix formulations and one coated-bead formulation. Reciprocation rate had the largest effect on time to 50% release for all matrix formulations. Top mesh size had an effect on release rate for those formulations having an erosion component to the drug release mechanism. Bottom mesh size had no effect on release rate. For the coated-bead formulation, no parameter had an effect on release rate. In an assessment of the hydrodynamics of the reciprocating cylinders, conditions equivalent to the 50 rpm paddle and 100 rpm basket were determined to be at the extreme low end of the available reciprocation rate range.

摘要

对美国药典溶出度测定仪3(往复圆筒)进行了评估,考察了仪器参数变化对六种亲水性基质制剂和一种包衣珠制剂药物释放速率的影响。往复速率对所有基质制剂达到50%释放时间的影响最大。顶部筛网尺寸对那些药物释放机制具有溶蚀成分的制剂的释放速率有影响。底部筛网尺寸对释放速率没有影响。对于包衣珠制剂,没有参数对释放速率有影响。在对往复圆筒的流体动力学评估中,确定相当于50转/分钟桨法和100转/分钟篮法的条件处于可用往复速率范围的极低端。