溶血对血糖仪测量的影响。

The effect of haemolysis on blood glucose meter measurement.

作者信息

Kilpatrick E S, Rumley A G, Rumley C N

机构信息

Department of Pathological Biochemistry, Gartnavel General Hospital, Glasgow, UK.

出版信息

Diabet Med. 1995 Apr;12(4):341-3. doi: 10.1111/j.1464-5491.1995.tb00489.x.

Abstract

A study was performed to assess the effect of varying degrees of sample haemolysis on the measurement of blood glucose by the Accutrend, Companion 2, ExacTech, Glucometer II, Glucometer 4, One Touch II, and Reflolux II blood glucose meters. Fresh venous blood was sonicated to induce complete haemolysis and then added in increasing proportions to homologous untreated blood to obtain nine samples with free haemoglobin concentrations up to 50 g l-1. The Accutrend meter showed the only significant (p < 0.05) linear relationship to degree of haemolysis (r = 0.988, p < 0.0001). For every 7% of red cells lysed, the Accutrend value increased by 15%. All other meters gave results which were within 15% of the non-haemolysed value. However, extreme (100%) haemolysis not only affected the Accutrend (glucose value 108% greater than reference) but also the ExacTech (+98%), the Glucometer II (-32%), and the Companion 2 (-41%). Thus, unwitting use of a haemolysed sample to measure whole blood glucose may, with the Accutrend in particular, lead to erroneous results.

摘要

进行了一项研究,以评估不同程度的样本溶血对Accutrend、Companion 2、ExacTech、血糖仪II、血糖仪4、One Touch II和Reflolux II血糖仪测量血糖的影响。将新鲜静脉血进行超声处理以诱导完全溶血,然后按递增比例添加到同源未处理血液中,以获得九个游离血红蛋白浓度高达50 g l-1的样本。Accutrend血糖仪显示出与溶血程度唯一显著(p < 0.05)的线性关系(r = 0.988,p < 0.0001)。每有7%的红细胞裂解,Accutrend测得的值就增加15%。所有其他血糖仪给出的结果与未溶血值相差在15%以内。然而,极端(100%)溶血不仅影响Accutrend(血糖值比参考值高108%),还影响ExacTech(+98%)、血糖仪II(-32%)和Companion 2(-41%)。因此,无意中使用溶血样本测量全血葡萄糖,特别是使用Accutrend血糖仪时,可能会导致错误结果。

文献AI研究员

20分钟写一篇综述,助力文献阅读效率提升50倍。

立即体验

用中文搜PubMed

大模型驱动的PubMed中文搜索引擎

马上搜索

文档翻译

学术文献翻译模型,支持多种主流文档格式。

立即体验