Douthwaite W, Pardhan S
Department of Optometry, University of Bradford, West Yorkshire, U.K.
Cornea. 1995 May;14(3):253-7. doi: 10.1097/00003226-199505000-00005.
The vertex (apical) radius and the p value of 30 conicoidal convex surface plastic buttons were measured using the Topcon KR-3500 autokeratometer. The vertex radius was found using the equations derived in the introduction. The results were compared to those derived by the Form Talysurf technique, which was considered to give a precise indication of both the vertex radius and p value. The results revealed a good agreement between the two methods of measurement. The autokeratometer displayed a tendency to underread the vertex radius, compared to the Talysurf result, but the maximum difference was 0.02 mm (95% confidence interval). For the p value, the autokeratometer could over- or underread with a maximum difference of 0.035. The largest coefficient of repeatability for the two observers using the Topcon autokeratometer was found to be 0.033 mm for vertex radius and 0.068 for the p value.
使用拓普康KR - 3500自动角膜曲率计测量了30个圆锥面凸面塑料纽扣的顶点(顶端)半径和p值。顶点半径通过引言中推导的公式得出。将结果与通过Form Talysurf技术得出的结果进行比较,该技术被认为能精确指示顶点半径和p值。结果表明两种测量方法之间具有良好的一致性。与Talysurf测量结果相比,自动角膜曲率计显示出低估顶点半径的趋势,但最大差异为0.02毫米(95%置信区间)。对于p值,自动角膜曲率计可能高估或低估,最大差异为0.035。使用拓普康自动角膜曲率计的两名观察者的最大重复性系数,顶点半径为0.033毫米,p值为0.068。