EyeSys角膜地形图测量应用于校准的椭球形凸面。

EyeSys corneal topography measurement applied to calibrated ellipsoidal convex surfaces.

作者信息

Douthwaite W A

机构信息

Department of Optometry, Bradford University, Bradford, West Yorks.

出版信息

Br J Ophthalmol. 1995 Sep;79(9):797-801. doi: 10.1136/bjo.79.9.797.

Abstract

AIMS/BACKGROUND: This study was carried out to assess the accuracy of the EyeSys videokeratoscope by using convex ellipsoidal surfaces of known form.

METHODS

PMMA convex ellipsoidal buttons were calibrated using Form Talysurf analysis which allowed subsequent calculation of the vertex radius and p value of the surface. The EyeSys videokeratoscope was used to examine the same ellipsoids. The tabular data provided by the instrument software were used to plot a graph of r2 versus y2 where r is the measured radius at y, the distance from the corneal point being measured to the surface vertex. The intercept on the ordinate of this graph gives the vertex radius and the slope the p value. The results arising from the Talysurf and the EyeSys techniques were compared.

RESULTS

The EyeSys videokeratoscope gave readings for both vertex radius and p value that were higher than those of the Talysurf analysis. The vertex radius was around 0.1 mm greater. The p value results were similar by the two methods for p values around unity but the EyeSys results were higher and the discrepancy increased as the p value approached that of a paraboloid.

CONCLUSIONS

Although the videokeratoscope may be useful in comparative studies of the cornea, there must be some doubt about the absolute values displayed. The disagreement is sufficiently large to suggest that the instrument may not be accurate enough for contact lens fitting purposes.

摘要

目的/背景:本研究旨在通过使用已知形状的凸椭圆形表面来评估EyeSys角膜地形图仪的准确性。

方法

使用Form Talysurf分析仪对聚甲基丙烯酸甲酯(PMMA)凸椭圆形纽扣进行校准,从而能够随后计算表面的顶点半径和p值。使用EyeSys角膜地形图仪检查相同的椭圆形。仪器软件提供的表格数据用于绘制r2与y2的关系图,其中r是在y处测量的半径,y是从被测量的角膜点到表面顶点的距离。该图纵坐标上的截距给出顶点半径,斜率给出p值。比较了Talysurf和EyeSys技术得出的结果。

结果

EyeSys角膜地形图仪给出的顶点半径和p值读数均高于Talysurf分析的结果。顶点半径大约大0.1毫米。对于p值接近1的情况,两种方法得到的p值结果相似,但EyeSys的结果更高,并且随着p值接近抛物面的p值,差异增大。

结论

尽管角膜地形图仪在角膜的比较研究中可能有用,但对于所显示的绝对值必须存在一些疑问。差异足够大,表明该仪器对于隐形眼镜验配目的可能不够准确。

相似文献

3
Comparison of videokeratoscope and autokeratometer measurements on ellipsoid surfaces and human corneas.
J Refract Surg. 1998 Jul-Aug;14(4):414-9. doi: 10.3928/1081-597X-19980701-07.
4
Corneal topography by keratometry.通过角膜曲率计进行角膜地形图检查。
Br J Ophthalmol. 2000 Aug;84(8):842-7. doi: 10.1136/bjo.84.8.842.

本文引用的文献

文献检索

告别复杂PubMed语法,用中文像聊天一样搜索,搜遍4000万医学文献。AI智能推荐,让科研检索更轻松。

立即免费搜索

文件翻译

保留排版,准确专业,支持PDF/Word/PPT等文件格式,支持 12+语言互译。

免费翻译文档

深度研究

AI帮你快速写综述,25分钟生成高质量综述,智能提取关键信息,辅助科研写作。

立即免费体验