Suppr超能文献

特发性成人脑积水综合征患者的脑脊液分流动力学

Cerebrospinal fluid shunt dynamics in patients with idiopathic adult hydrocephalus syndrome.

作者信息

Malm J, Kristensen B, Fagerlund M, Koskinen L O, Ekstedt J

机构信息

Department of Neurology, University Hospital of Northern Sweden, Umeå.

出版信息

J Neurol Neurosurg Psychiatry. 1995 Jun;58(6):715-23. doi: 10.1136/jnnp.58.6.715.

Abstract

The objective was to assess CSF dynamics of different shunt constructions in patients with adult hydrocephalus syndrome and correlate these findings to clinical outcome, neuroradiology, and the specifications of the shunts provided by the manufacturer. Thirty four patients with idiopathic adult hydrocephalus (normal pressure hydrocephalus) syndrome were included in a prospective, consecutive case series. A differential pressure valve (Cordis Hakim standard system) was used in 28 patients and a variable resistance valve (Cordis Orbis-Sigma) in six. A constant pressure infusion method was used; CSF pressure and conductance were determined before surgery. Three months after shunt placement CSF pressure, the "pressure v flow" curve, and gravity induced flow were measured. There was no difference between mean preoperative and postoperative resting CSF pressures in patients with Hakim shunts. The opening pressures of the Hakim shunts were higher than the value proposed by the manufacturer. A pronounced gravity effect induced CSF flow and decrease of the CSF pressure. In functioning variable resistance valves, CSF dynamics normalised postoperatively. There was no gravity effect and the characteristics shaped "pressure v flow" curve was sometimes seen. Six patients (three differential pressure valves, three variable resistance valves) had non-functioning shunts. Four of these patients were improved after the operation but improvement was transient in three. In all patients, there was no relation between the width of the ventricles and clinical improvement or CSF pressure. In conclusion, the differential pressure valve system does not behave according to the specifications provided by the manufacturer. A decrease in CSF pressure in patients with this shunt was solely due to the effect of gravity. Eleven percent of the differential pressure valves and 50% of the variable resistance valves were non-functioning. In the functioning variable resistance valves, the antisiphon system seems to be effective.

摘要

目的是评估成人脑积水综合征患者不同分流装置的脑脊液动力学,并将这些结果与临床结局、神经放射学以及制造商提供的分流装置规格相关联。34例特发性成人脑积水(正常压力脑积水)综合征患者纳入前瞻性连续病例系列。28例患者使用了压差阀(Cordis Hakim标准系统),6例使用了可变电阻阀(Cordis Orbis-Sigma)。采用恒压输注法;术前测定脑脊液压力和传导率。分流放置3个月后,测量脑脊液压力、“压力-流量”曲线和重力诱导流量。使用Hakim分流装置的患者术前和术后静息脑脊液压力平均值无差异。Hakim分流装置的开启压力高于制造商建议的值。明显的重力效应诱导脑脊液流动并降低脑脊液压力。在功能正常的可变电阻阀中,脑脊液动力学术后恢复正常。没有重力效应,有时可见“压力-流量”曲线的特征形状。6例患者(3例压差阀,3例可变电阻阀)分流装置功能异常。其中4例患者术后有所改善,但3例改善是短暂的。所有患者中,脑室宽度与临床改善或脑脊液压力之间均无关联。总之,压差阀系统的表现不符合制造商提供的规格。使用这种分流装置的患者脑脊液压力降低完全是由于重力作用。11%的压差阀和50%的可变电阻阀功能异常。在功能正常的可变电阻阀中,抗虹吸系统似乎有效。

https://cdn.ncbi.nlm.nih.gov/pmc/blobs/b032/1073553/594ad04e564c/jnnpsyc00030-0068-a.jpg

文献检索

告别复杂PubMed语法,用中文像聊天一样搜索,搜遍4000万医学文献。AI智能推荐,让科研检索更轻松。

立即免费搜索

文件翻译

保留排版,准确专业,支持PDF/Word/PPT等文件格式,支持 12+语言互译。

免费翻译文档

深度研究

AI帮你快速写综述,25分钟生成高质量综述,智能提取关键信息,辅助科研写作。

立即免费体验