一种用于无创测定 Björk-Shiley 二尖瓣压力梯度的超声多普勒技术。

An ultrasound Doppler technique for the noninvasive determination of the pressure gradient in the Björk-Shiley mitral valve.

作者信息

Holen J, Simonsen S, Frøysaker T

出版信息

Circulation. 1979 Mar;59(3):436-42. doi: 10.1161/01.cir.59.3.436.

Abstract

The accuracy in determining the pressure gradient in the Björk-Shiley mitral implant from noninvasive ultrasound Doppler data was explored in nine adult patients. Manometric and ultrasound data were collectd simultaneously, and identical diastolic periods were used to compare the manometric gradient (delta Pm) with the gradient obtained from ultrasound data (delta Pu). In the nine patients the mean diastolic value of delta Pm ranged from 2-12.5 mm Hg and the difference between the mean diastolic values of delta Pm and delta Pu was 0.3 +/- 1.0 mm Hg(SD). The results of the investigation indicated that the method is accurate and reliable in the nonivasive determination of the mean diastolic gradient in the Björk-Shiley mitral implant.

摘要

对9名成年患者进行了研究,探讨了根据无创超声多普勒数据确定 Björk-Shiley 二尖瓣植入物压力梯度的准确性。同时收集了测压数据和超声数据,并使用相同的舒张期来比较测压梯度(δPm)和从超声数据获得的梯度(δPu)。9名患者的δPm舒张期平均值在2至12.5 mmHg之间,δPm和δPu舒张期平均值之间的差异为0.3±1.0 mmHg(标准差)。研究结果表明,该方法在无创测定 Björk-Shiley 二尖瓣植入物舒张期平均梯度方面准确可靠。