Measurements and calculations of the absorbed dose distribution around a 60Co source.

作者信息

Tiourina T B, Dries W J, van der Linden P M

机构信息

Department of Radiotherapy, Catharina Hospital, Eindhoven, The Netherlands.

出版信息

Med Phys. 1995 May;22(5):549-54. doi: 10.1118/1.597583.

Abstract

The data from Meisberger et al. [Radiology 90, 953-957 (1968)] are often used as a basis for dose calculations in brachytherapy. In order to describe the absorbed dose in water around a brachytherapy point source, Meisberger provided a polynomial fit for different isotopes taking into account the effect of attenuation and scattering. The validity of the Meisberger coefficients is restricted to distances up to 10 cm from the source, which is regarded to be satisfactory for most brachytherapy applications. However, for more distant organs it may lead to errors in calculated absorbed dose. For this reason dose measurements have been performed in air and in water around a high activity 60Co source used in high dose rate brachytherapy. Measurements were carried out to distances of 20 cm, using ionization chambers. These data show that at a distance of about 15 cm the amount of scattered radiation virtually equals the amount of primary radiation. This emphasizes the contribution of scattered radiation to the dose in healthy tissue far from the target volume, even with relatively high energy photon radiation of 60Co. It is also shown that the Meisberger data as well as the approach of Van Kleffens and Star [Int. J. Radiat. Oncol. Phys. 5, 557-563 (1979)] lead to significant errors in absorbed dose between distances of 10 and 20 cm from the source. In addition to these measurements, the Monte Carlo code has been used to calculate separately primary dose and scattered dose from a cobalt point source. The calculated results agree with the experimental data within 1% for a most distant dose scoring region.

摘要

文献AI研究员

20分钟写一篇综述,助力文献阅读效率提升50倍。

立即体验

用中文搜PubMed

大模型驱动的PubMed中文搜索引擎

马上搜索

文档翻译

学术文献翻译模型,支持多种主流文档格式。

立即体验