Kojima K Y, Szumowski J, Sheley R C, Quinn S F
Department of Radiology, Good Samaritan Hospital and Medical Center, Portland, Ore, USA.
Radiology. 1995 Sep;196(3):871-5. doi: 10.1148/radiology.196.3.7644660.
In a comparison of conventional and magnetic resonance (MR) angiography of the lower extremities, MR imaging was performed with extremity and body coils and a unilateral six-coil telescopic phased-array coil. In one of seven volunteers, average signal-to-noise ratio was 140% higher with the latter (p < .01), only three of five stations were necessary, and examination time was decreased 50%. In the four patients, correlation in findings was 100%.
在一项下肢传统血管造影与磁共振(MR)血管造影的对比研究中,采用肢体线圈、体部线圈及单侧六线圈可伸缩相控阵线圈进行MR成像。在7名志愿者中的1名身上,使用后者时平均信噪比提高了140%(p < 0.01),仅需对5个部位中的3个进行检查,且检查时间缩短了50%。在4名患者中,检查结果的相关性为100%。