Asymmetric fan transmission CT on SPECT systems.

作者信息

Chang W, Loncaric S, Huang G, Sanpitak P

机构信息

Rush Presbyterian-St Luke's Medical Center, Chicago, IL 60612, USA.

出版信息

Phys Med Biol. 1995 May;40(5):913-28. doi: 10.1088/0031-9155/40/5/013.

Abstract

For proper attenuation correction of SPECT images, a set of 3D attenuation maps specific to the imaging slices is needed. Among the many different approaches for deriving the attenuation maps, fan beam transmission CT (FBTCT), performed on the same SPECT system as emission imaging, has many promising and clinically practical features. The major problem of FBTCT is that the current SPECT systems do not have a large enough field of view (FOV) to cover the typical cross-sectional size of patients. To address this problem, we have developed a novel asymmetric fan (AsF) sampling scheme to extend the FOV to practical sizes for clinical TCT imaging on existing SPECT systems. This AsF scheme samples only half of the intended FOV in each projection; the other half would be sampled in an opposing projection after detector rotation. We have implemented the AsF sampling on a three-head SPECT system through a specially designed source-collimator assembly. We have modified the conventional convolution backprojection algorithm to facilitate simple and fast image reconstruction. The feasibility of the approach is confirmed by the quality of the derived TCT images of various phantoms and human subjects. The AsF sampling scheme could also have applications in other general transmission CT systems.

摘要

文献检索

告别复杂PubMed语法,用中文像聊天一样搜索,搜遍4000万医学文献。AI智能推荐,让科研检索更轻松。

立即免费搜索

文件翻译

保留排版,准确专业,支持PDF/Word/PPT等文件格式,支持 12+语言互译。

免费翻译文档

深度研究

AI帮你快速写综述,25分钟生成高质量综述,智能提取关键信息,辅助科研写作。

立即免费体验