Kelly D J, Chan C T, Paxton H, Thompson K, Howard R, Dasch G A
Rickettsial Diseases Program, Naval Medical Research Institute, Bethesda, Maryland 20889-5055, USA.
Clin Diagn Lab Immunol. 1995 May;2(3):356-60. doi: 10.1128/cdli.2.3.356-360.1995.
A commercial enzyme immunoassay kit called the Dip-S-Ticks (DS) for the detection of total immunoglobulin (Ig) G and IgM human antibodies to Rickettsia typhi was evaluated. In tests with 340 serum samples from patients with diagnosed cases of rickettsial diseases, patients suffering from other febrile illnesses, and normal subjects, the DS compared favorably with the standard indirect fluorescent-antibody (IFA) test. At IFA cutoff titers of > or = 1.64 and > or = 1:128, the DS showed sensitivities of 88.2 and 91.4% and specificities of 91.8 and 87.7%, respectively. The DS test correlated significantly with both the IFA IgG (r = 0.84, P < 0.0005) and IgM (r = 0.63, P < 0.0005) titers. Only 80% of IgG and 82% of IgM IFA readings determined by two technicians were within one dilution, while the DS was more reliable, with 100% within one dot. The rapidity, reliability, and simplicity of the DS suggest that it is a suitable test for use in clinical laboratories unable to perform the IFA test.
对一种名为Dip-S-Ticks(DS)的用于检测人抗伤寒立克次体总免疫球蛋白(Ig)G和IgM抗体的商用酶免疫分析试剂盒进行了评估。在对340份血清样本进行检测时,这些样本来自确诊立克次体病的患者、患有其他发热性疾病的患者以及正常受试者,DS与标准间接荧光抗体(IFA)检测相比表现良好。在IFA临界滴度分别为≥1.64和≥1:128时,DS的灵敏度分别为88.2%和91.4%,特异性分别为91.8%和87.7%。DS检测与IFA IgG(r = 0.84,P < 0.0005)和IgM(r = 0.63,P < 0.0005)滴度均显著相关。两名技术人员测定的IFA IgG读数中只有80%以及IgM读数中只有82%在一个稀释度范围内,而DS更可靠,100%在一个点内。DS的快速、可靠和简便表明它是一种适合在无法进行IFA检测的临床实验室中使用的检测方法。