Suppr超能文献

用于在关节镜控制下测量软骨硬度的压痕器械。

Indentation instrument for the measurement of cartilage stiffness under arthroscopic control.

作者信息

Lyyra T, Jurvelin J, Pitkänen P, Väätäinen U, Kiviranta I

机构信息

Department of Anatomy, University of Kuopio, Finland.

出版信息

Med Eng Phys. 1995 Jul;17(5):395-9. doi: 10.1016/1350-4533(95)97322-g.

Abstract

Changes in the biomechanical properties of articular cartilage are one of the first signs of the tissue degeneration. We have developed a small size indentation instrument for the quantification of cartilage stiffness under arthroscopic control. During measurement, the indenter imposes a constant deformation on the cartilage and the maximal indenter force, by which the cartilage resists the deformation, is used as a measure for cartilage stiffness. The instrument has been tested in laboratory conditions with elastomer and cadaver knee joint cartilage samples. A linear relationship was found between indenter force and elastomer stiffness (r = 0.990, n = 14) as well as between indenter force and cartilage shear modulus obtained by a reference device (r = 0.879, n = 22). Also, the correlation between two repeated measurements at the measurement sites, used to evaluate the reproducibility, was linear (r = 0.953, n = 16). Quantitative detection of cartilage stiffness is possible with the instrument.

摘要

关节软骨生物力学特性的改变是组织退变的最早迹象之一。我们开发了一种小型压痕仪,用于在关节镜控制下定量评估软骨硬度。在测量过程中,压头对软骨施加恒定变形,软骨抵抗变形的最大压头力用作软骨硬度的度量。该仪器已在实验室条件下用弹性体和尸体膝关节软骨样本进行了测试。发现压头力与弹性体硬度之间存在线性关系(r = 0.990,n = 14),以及压头力与通过参考设备获得的软骨剪切模量之间存在线性关系(r = 0.879,n = 22)。此外,用于评估可重复性的测量部位两次重复测量之间的相关性也是线性的(r = 0.953,n = 16)。使用该仪器可以对软骨硬度进行定量检测。

文献检索

告别复杂PubMed语法,用中文像聊天一样搜索,搜遍4000万医学文献。AI智能推荐,让科研检索更轻松。

立即免费搜索

文件翻译

保留排版,准确专业,支持PDF/Word/PPT等文件格式,支持 12+语言互译。

免费翻译文档

深度研究

AI帮你快速写综述,25分钟生成高质量综述,智能提取关键信息,辅助科研写作。

立即免费体验