Roberts D H, Baxter S E, Brennan P T, Gammage M D
Department of Cardiology, Birmingham General Hospital, United Kingdom.
Pacing Clin Electrophysiol. 1995 Jan;18(1 Pt 1):65-9. doi: 10.1111/j.1540-8159.1995.tb02477.x.
The ability of externally strapped accelerometer- (Excel [Cardiac Pacemakers, Inc.]) and vibration-based (Activitrax [Medtronic, Inc.]) rate adaptive pacemakers to reproduce the rate response of the same implanted devices with identical programming was evaluated in ten patients by ambulatory Holter monitoring. The resting and postexercise external pacemaker rates closely resembled those of the respective implanted devices. During short bursts and more prolonged exercise, both types of strapped-on devices underestimated maximal implanted pacemaker rate response by 4%-10% when programmed to nominal rate adaptive settings. Studies evaluating chronotropic responses from either type of externally strapped activity sensor appear valid, provided the modest attenuation in maximal rate increase by this method is appreciated.
通过动态心电图监测,对10例患者评估了外部佩戴的加速度计式(Excel[心脏起搏器公司])和基于振动式(Activitrax[美敦力公司])频率适应性起搏器在相同程控条件下重现同一植入式设备频率反应的能力。静息和运动后外部起搏器频率与各自植入式设备的频率非常相似。在短时间爆发性运动和较长时间运动期间,当程控为标称频率适应性设置时,两种类型的外部佩戴设备将植入式起搏器的最大频率反应低估了4%-10%。只要认识到通过这种方法最大心率增加幅度会有适度衰减,那么评估任何一种外部佩戴活动传感器的变时反应的研究似乎都是有效的。