Ogawa K
Department of Electronic Informatics, College of Engineering, Hosei University, Japan.
Ann Nucl Med. 1994 Nov;8(4):277-81. doi: 10.1007/BF03165031.
In a quantitative SPECT study with multiple radionuclides, it is very important to eliminate the counts of scattered photons from planar images. In this paper, the triple energy window (TEW) method, which has developed to eliminate the counts of scatter photons in measured counts, was applied to a multiradionuclide SPECT study and its effect was examined in a simulation study. In the simulation, we used Tc-99m and Tl-201, and we assumed their photopeak energies to be 141 and 73 keV, respectively. For two different activity distributions in a cylinder phantom, simulation tests with Tl-201 and Tc-99m gave good agreement between the activity distributions reconstructed from primary photons and those from corrected data. The contrast of a cold spot area in images with and without correction improved around 70% to more than 96%.
在一项使用多种放射性核素的定量单光子发射计算机断层扫描(SPECT)研究中,从平面图像中消除散射光子的计数非常重要。本文将已开发用于消除测量计数中散射光子计数的三能量窗(TEW)方法应用于多放射性核素SPECT研究,并在模拟研究中检验了其效果。在模拟中,我们使用了锝-99m(Tc-99m)和铊-201(Tl-201),并假设它们的光电峰能量分别为141 keV和73 keV。对于圆柱模型中两种不同的活度分布,使用Tl-201和Tc-99m进行的模拟测试表明,从一次光子重建的活度分布与校正后数据重建的活度分布之间具有良好的一致性。校正前后图像中冷区的对比度提高了约70%至96%以上。