SPECT dual-energy-window Compton correction: scatter multiplier required for quantification.

作者信息

Koral K F, Swailem F M, Buchbinder S, Clinthorne N H, Rogers W L, Tsui B M

机构信息

Department of Internal Medicine, University of Michigan Medical Center, Ann Arbor 48109-0552.

出版信息

J Nucl Med. 1990 Jan;31(1):90-8.

PMID:2404095
Abstract

The dual-energy window Compton-scattering correction technique is defined here especially for accurate quantification of focal regions having higher than average uptake. The quantification is relative to a known-activity reference source. The scatter multiplier ("k" value) is determined for a radioactive 99mTc sphere on or off the axis of a cylinder containing water with or without background. Both maximum likelihood and filtered-backprojection reconstruction are employed. Either projections or tomograms are corrected. With tight regions of interest, there is a tendency for the requisite "k" value to be slightly lower as the diameter of the cylinder is increased. Neither sphere location nor background perturbs "k", however, so a constant value is a good, first approximation. Then a two-sphere validation test yields an accuracy of 8% with subtracted-tomograms ("k" = 1.30) and 2% with subtracted-projections ("k" = 1.20). With a reference-source region of interest which is four times larger, "k" is reduced and also now depends on background. Although equivalent quantitatively, maximum likelihood is preferable to filtered backprojection with Chang attenuation correction since it produces a less-noisy image.

摘要

文献检索

告别复杂PubMed语法,用中文像聊天一样搜索,搜遍4000万医学文献。AI智能推荐,让科研检索更轻松。

立即免费搜索

文件翻译

保留排版,准确专业,支持PDF/Word/PPT等文件格式,支持 12+语言互译。

免费翻译文档

深度研究

AI帮你快速写综述,25分钟生成高质量综述,智能提取关键信息,辅助科研写作。

立即免费体验