美国放射肿瘤学组(RTOG)关于超声热疗温度测量指南的建议修订版。

A recommended revision in the RTOG thermometry guidelines for hyperthermia administered by ultrasound.

作者信息

Waterman F M, Hoh L L

机构信息

Department of Radiation Oncology and Nuclear Medicine, Thomas Jefferson University Hospital, Philadelphia, PA 19107.

出版信息

Int J Hyperthermia. 1995 Jan-Feb;11(1):121-30. doi: 10.3109/02656739509004953.

Abstract

RTOG thermometry guidelines for clinical trials of hyperthermia using planar ultrasound recommended that temperatures be mapped in polyurethane catheters by use of single-junction copper-constantan thermocouples. These guidelines were based on an assumption that the error in temperature measurement due to thermal conduction would generally not exceed +/- 0.3 degrees C. The validity of this assumption was tested with a commercially available single-junction copper-constantan thermocouple. The width of the point spread function, an indicator of the relative magnitude of the conduction error, was five times greater than expected. As a result, the conduction error is projected to exceed 0.3 degrees C in a temperature gradient of only 1.5 degrees C/cm. This projection was confirmed by mapping a thermal peak which simulates a typical clinical temperature profile. This peak had an amplitude of 6 degrees C, a full-width at half-maximum of 3.5 cm, and a maximum gradient of approximately 3 degrees C/cm. Temperatures measured at 0.5-cm intervals over the span of this peak were in error by a mean of +/- 0.6 degrees C. It is strongly recommended that the RTOG guidelines be revised to replace copper-constantan thermocouples with manganin-constantan single- or multi-junction thermocouples which will assure that the conduction error will be < +/- 0.3 degrees C.

摘要

放射肿瘤学组(RTOG)关于使用平面超声进行热疗临床试验的温度测量指南建议,应使用单结铜-康铜热电偶在聚氨酯导管中绘制温度图。这些指南基于这样一个假设,即由于热传导导致的温度测量误差通常不会超过±0.3摄氏度。使用市售的单结铜-康铜热电偶对这一假设的有效性进行了测试。点扩散函数的宽度是传导误差相对大小的一个指标,比预期大五倍。因此,在仅1.5摄氏度/厘米的温度梯度下,传导误差预计将超过0.3摄氏度。通过绘制模拟典型临床温度分布的热峰证实了这一预测。该峰的幅度为6摄氏度,半高宽为3.5厘米,最大梯度约为3摄氏度/厘米。在该峰的跨度上以0.5厘米的间隔测量的温度平均误差为±0.6摄氏度。强烈建议修订RTOG指南,用锰铜-康铜单结或多结热电偶取代铜-康铜热电偶,以确保传导误差将<±0.3摄氏度。

文献AI研究员

20分钟写一篇综述,助力文献阅读效率提升50倍。

立即体验

用中文搜PubMed

大模型驱动的PubMed中文搜索引擎

马上搜索

文档翻译

学术文献翻译模型,支持多种主流文档格式。

立即体验