在为铜-康铜热电偶设计的测温装置中使用锰铜-康铜热电偶。

Use of manganin-constantan thermocouples in thermometry units designed for copper-constantan thermocouples.

作者信息

Hoh L L, Waterman F M

机构信息

Department of Radiation Oncology and Nuclear Medicine, Thomas Jefferson University Hospital, Philadelphia, PA 19107.

出版信息

Int J Hyperthermia. 1995 Jan-Feb;11(1):131-8. doi: 10.3109/02656739509004954.

Abstract

Commercial ultrasound hyperthermia systems typically include thermometry units designed for copper-constantan thermocouples. Replacing these copper-constantan thermocouples with manganin-constantan thermocouples is advantageous in reducing the measurement error caused by the conduction of heat along the copper wire, but their performance in these thermometry units is uncertain. The accuracy of manganin-constantan thermocouples in the Labthermics LT-100, Clini-Therm TS1200/TM100, and Physitemp TM-12 thermometry units was investigated using a temperature controlled circulating water bath monitored by a mercury thermometer having a calibration traceable to NIST. The results demonstrate that an accuracy of +/- 0.2 degrees C can be achieved with manganin-constantan thermocouples over the range 35-55 degrees C without hardware modification provided specific calibration procedures are followed. With the Labthermics LT-100, a double point calibration should be carried out at 35 and 55 degrees C. With the Clini-Therm TS1200/TM100, a self-calibration of the unit using its internal calibration well plus a single point calibration using an external temperature standard provides sufficient accuracy. The Physitemp TM-12 requires an external computer for read out and the user must provide additional software to correct for the error by either a single or multiple point calibration.

摘要

商用超声热疗系统通常包括为铜-康铜热电偶设计的温度测量单元。用锰铜-康铜热电偶取代这些铜-康铜热电偶,有利于减少沿铜线热传导引起的测量误差,但其在这些温度测量单元中的性能尚不确定。使用由可溯源至美国国家标准与技术研究院(NIST)校准的汞温度计监测的温度控制循环水浴,研究了锰铜-康铜热电偶在Labthermics LT-100、Clini-Therm TS1200/TM100和Physitemp TM-12温度测量单元中的精度。结果表明,如果遵循特定的校准程序,在35至55摄氏度范围内,使用锰铜-康铜热电偶无需硬件修改即可实现±0.2摄氏度的精度。对于Labthermics LT-100,应在35和55摄氏度下进行双点校准。对于Clini-Therm TS1200/TM100,使用其内部校准井对单元进行自校准,再加上使用外部温度标准进行单点校准,可提供足够的精度。Physitemp TM-12需要外部计算机进行读数,用户必须提供额外的软件,通过单点或多点校准来校正误差。

相似文献

3
RTOG quality assurance guidelines for clinical trials using hyperthermia administered by ultrasound.
Int J Radiat Oncol Biol Phys. 1991 May;20(5):1099-107. doi: 10.1016/0360-3016(91)90211-l.
4
7
Hyperthermia thermometry evaluation: criteria and guidelines.
Int J Radiat Oncol Biol Phys. 1988 Feb;14(2):327-35. doi: 10.1016/0360-3016(88)90440-3.

文献AI研究员

20分钟写一篇综述,助力文献阅读效率提升50倍。

立即体验

用中文搜PubMed

大模型驱动的PubMed中文搜索引擎

马上搜索

文档翻译

学术文献翻译模型,支持多种主流文档格式。

立即体验