The usefulness of the Noise-Field Test as a screening method for visual field defects.

作者信息

Adachi M, Shirato S

机构信息

Department of Ophthalmology, University of Tokyo, School of Medicine, Japan.

出版信息

Jpn J Ophthalmol. 1994;38(4):392-9.

PMID:7723208
Abstract

In 1990, we reported the usefulness of the noise-field spontaneously generated on a home television screen for the subjective perception of visual field defects (VFD) in glaucoma patients (Noise-Field Test: NFT). Recently, the sensitivity and specificity of the NFT for testing 555 eyes of primary open angle glaucoma patients and 300 eyes of subjects with normal visual field were 89.2% and 93.7%, respectively. We also performed a comparative study among the NFT and Tübingen electric campimetry (TEC) and oculokinetic perimetry (OKP), and also evaluated the efficiency of the NFT using participants examined in general health check-up systems. In the comparative study testing 136 eyes of glaucoma patients who had never perceived their VFD, and 114 eyes of subjects with normal visual field, the sensitivity of the NFT, TEC (fine), TEC (coarse), and OKP was 85.3%, 75.0%, 71.3% or 59.5%, respectively. The specificity was 88.6%, 78.9%, 82.5% or 95.6%, respectively. In the study made in the general health check-up system, the NFT was carried out on 935 participants (1870 eyes). After the NFT, all participants received an ophthalmological examination. Those participants whose fundus was suspected to have VFD as well as participants who reported abnormalities of the noise-field were examined by static perimetry. One hundred and sixty-two eyes perceived abnormalities of the noise-field; 58 eyes were judged as having VFD by an ophthalmologist and the defects were confirmed.(ABSTRACT TRUNCATED AT 250 WORDS)

摘要

文献检索

告别复杂PubMed语法,用中文像聊天一样搜索,搜遍4000万医学文献。AI智能推荐,让科研检索更轻松。

立即免费搜索

文件翻译

保留排版,准确专业,支持PDF/Word/PPT等文件格式,支持 12+语言互译。

免费翻译文档

深度研究

AI帮你快速写综述,25分钟生成高质量综述,智能提取关键信息,辅助科研写作。

立即免费体验