Matsumoto M, Horikoshi H, Moteki T, Hatori N, Tateno Y, Iinuma T, Matsumoto T, Yamamoto S, Baba T
Department of Diagnostic Radiology, Gunma University Hospital.
Nihon Igaku Hoshasen Gakkai Zasshi. 1995 Feb;55(3):172-9.
We have developed computed tomography (CT) equipment for lung-cancer screening (named LSCT) that can be used exclusively for lung-cancer screening with spiral volumetric CT and is available on a screening car. A pilot study with LSCT was performed from November 1992 to January 1993 on 118 screenees at the secondary examination of lung-cancer screening. Scan parameters were as follows: 120 kVp, 50 mA, slice thickness 10 mm, table feed 10 mm/sec, scan time 2 sec/rotation. All the screenees were scanned under quiet respiration instead of the breath-hold technique. Under these scan parameters, LSCT images were almost free from respiratory motion artifacts even at the lung base. Continuity of the bronchial tree and vessels was well maintained in consecutive slices. Pulmonary nodules approximately 5 mm in diameter were clearly depicted. By LSCT, 43 of 118 screenees were diagnosed to need further examinations. And 33 out of 43 screenees underwent detailed examinations. Finally, 16 lung cancers were confirmed. Ten of 16 patients with lung cancer underwent surgery; nine were in stage I and one in stage IIIA. LSCT was considered to be useful in lung-cancer screening.
我们已开发出用于肺癌筛查的计算机断层扫描(CT)设备(名为LSCT),它可专门用于螺旋容积CT肺癌筛查,且可在筛查车上使用。1992年11月至1993年1月,对118名肺癌筛查二次检查的受检者进行了LSCT试点研究。扫描参数如下:120 kVp、50 mA、层厚10 mm、床速10 mm/秒、扫描时间2秒/旋转。所有受检者均在平静呼吸下扫描,而非屏气技术。在这些扫描参数下,即使在肺底部,LSCT图像也几乎没有呼吸运动伪影。连续层面中支气管树和血管的连续性保持良好。直径约5 mm的肺结节清晰显示。通过LSCT,118名受检者中有43人被诊断需要进一步检查。43名受检者中有33人接受了详细检查。最终,确诊16例肺癌。16例肺癌患者中有10例接受了手术;9例为I期,1例为IIIA期。LSCT被认为对肺癌筛查有用。