在乳腺摄影中使用对比度-细节测试模体优化光学密度。

The use of a contrast-detail test object in the optimization of optical density in mammography.

作者信息

Robson K J, Kotre C J, Faulkner K

机构信息

Regional Medical Physics Department, Newcastle General Hospital, Newcastle upon Tyne, UK.

出版信息

Br J Radiol. 1995 Mar;68(807):277-82. doi: 10.1259/0007-1285-68-807-277.

Abstract

In mammography, it is important that the maximum amount of diagnostic information is obtained from each radiograph. One of the factors influencing image quality is the optical density of the film. In this study, a contrast-detail test object was used to establish the optimum optical density value on the basis of signal-to-noise ratio (SNR) for two mammographic film-screen combinations under particular processing conditions. The optimum optical density, taken as being the point at which the threshold contrast is a minimum varied little with detail size. The optimum optical density for the Fuji film-screen combination with Photosol chemistry was found to be higher than that for the Kodak film-screen combination and Kodak processing conditions studied. Although the actual values of optical density are specific to the fixed processing conditions encountered, the technique has more general applicability.

摘要

在乳腺X线摄影中,从每张射线照片中获取最大量的诊断信息非常重要。影响图像质量的因素之一是胶片的光学密度。在本研究中,使用了一个对比度细节测试物体,根据特定处理条件下两种乳腺X线胶片-增感屏组合的信噪比(SNR)来确定最佳光学密度值。最佳光学密度被视为阈值对比度最小的点,随细节尺寸变化不大。发现采用Photosol化学处理的富士胶片-增感屏组合的最佳光学密度高于所研究的柯达胶片-增感屏组合及柯达处理条件下的最佳光学密度。尽管光学密度的实际值特定于所遇到的固定处理条件,但该技术具有更广泛的适用性。