Czerw R J, Fulkerson M S, Donnelly J C
Department of Endodontics, U.S. Army Dental Activity, Ft. Hood, TX 76544, USA.
J Endod. 1994 Dec;20(12):605-6. doi: 10.1016/S0099-2399(06)80086-5.
This article describes an in vitro test of the accuracy of a model for demonstrating the operation of electronic canal length measuring devices (ECLMDs). The model is made from commercially available gelatin which is mixed with sterile saline solution. Extracted human canine and anterior teeth with fully formed apices had access preparations and complete pulpectomies before drying the canals with paper points. The tooth length was determined with an Exact-A-Pex using the gelatin test model. When the apex locator green light-emitting diode indicated 1.5 mm from the apex, the file length to the nearest 0.5 mm was measured with a rubber stop. This length was compared with the length determined by visualization of the tip of a #10 K-Flex file at the foramen. This method was repeated with the Foramatron IV using the instrument's digital reading of zero as the apex location and root canal length. There was complete agreement as to canal length between electronic root canal measuring devices and visual observations. This indicates that the gelatin test model is an accurate in vitro method of determining canal length.
本文描述了一种体外测试模型准确性的方法,该模型用于演示电子根管长度测量仪(ECLMDs)的操作。该模型由市售明胶与无菌盐溶液混合制成。将已萌出的具有完全形成根尖的人类犬齿和前牙进行开髓预备并完成牙髓摘除术,然后用纸尖干燥根管。使用明胶测试模型,通过Exact - A - Pex确定牙齿长度。当根尖定位仪的绿色发光二极管显示距离根尖1.5毫米时,用橡胶止动器测量锉的长度,精确到最接近的0.5毫米。将该长度与通过观察10号K - Flex锉尖在根尖孔处确定的长度进行比较。使用Foramatron IV重复该方法,将仪器的数字读数零作为根尖位置和根管长度。电子根管测量装置与视觉观察在根管长度方面完全一致。这表明明胶测试模型是一种准确的体外确定根管长度的方法。