乳牙根管长度测定中两种电子根尖定位仪准确性的体外评估

Ex vivo evaluation of the accuracy of two electronic apex locators during root canal length determination in primary teeth.

作者信息

Leonardo M R, Silva L A B, Nelson-Filho P, Silva R A B, Raffaini M S G G

机构信息

Department of Pediatric, Preventive and Community Dentistry, Ribeirão Preto School of Dentistry, University of São Paulo, Ribeirão Preto, São Paulo, Brazil.

出版信息

Int Endod J. 2008 Apr;41(4):317-21. doi: 10.1111/j.1365-2591.2007.01366.x. Epub 2008 Jan 19.

Abstract

AIM

To evaluate ex vivo the accuracy of two electronic apex locators during root canal length determination in primary incisor and molar teeth with different stages of physiological root resorption.

METHODOLOGY

One calibrated examiner determined the root canal length in 17 primary incisors and 16 primary molars (total of 57 root canals) with different stages of root resorption based on the actual canal length and using two electronic apex locators. Root canal length was measured both visually, with the placement of a K-file 1 mm short of the apical foramen or the apical resorption bevel, and electronically using two electronic apex locators (Root ZX II--J. Morita Corp. and Mini Apex Locator--SybronEndo) according to the manufacturers' instructions. Data were analysed statistically using the intraclass correlation (ICC) test.

RESULTS

Comparison of the actual root canal length and the electronic root canal length measurements revealed high correlation (ICC = 0.99), regardless of the tooth type (single-rooted and multi-rooted teeth) or the presence/absence of physiological root resorption.

CONCLUSIONS

Root ZX II and Mini Apex Locator proved useful and accurate for apex foramen location during root canal length measurement in primary incisors and molars.

摘要

目的

在体外评估两种电子根尖定位仪在不同生理性牙根吸收阶段的乳切牙和乳磨牙根管长度测定中的准确性。

方法

一名经过校准的检查者根据实际根管长度,使用两种电子根尖定位仪,测定17颗乳切牙和16颗乳磨牙(共57个根管)在不同牙根吸收阶段的根管长度。通过将K锉放置在距根尖孔或根尖吸收斜面1mm处进行视觉测量根管长度,并根据制造商的说明使用两种电子根尖定位仪(Root ZX II——日本森田公司和Mini Apex Locator——SybronEndo)进行电子测量。使用组内相关系数(ICC)检验对数据进行统计学分析。

结果

实际根管长度与电子根管长度测量结果的比较显示出高度相关性(ICC = 0.99),无论牙齿类型(单根牙和多根牙)或是否存在生理性牙根吸收。

结论

在乳切牙和乳磨牙根管长度测量中,Root ZX II和Mini Apex Locator被证明在根尖孔定位方面是有用且准确的。