Palmer B M, Sajjad M, Rottenberg D A
PET Imaging Service, VA Medical Center, Minneapolis, MN 55417, USA.
Nucl Med Biol. 1995 Feb;22(2):241-9. doi: 10.1016/0969-8051(94)00085-x.
A computer controlled [15O]H2O injection system has been designed for multiple bolus injections during PET cerebral blood flow studies. The system is housed in a lead safe, and radiation exposure to nearby personnel is low; we have measured rates of approx. 100 microSv/h (10 mR/h) during 2 min of [15O]H2O production and < 50 microSv/h (< 5 mR/h) after 3.7 GBq (100 mCi) has been produced. Reproducibility studies show that the injected activity varies with a coefficient of variation of 2-3%.
已设计出一种计算机控制的[15O]H2O注射系统,用于正电子发射断层扫描(PET)脑血流研究中的多次团注注射。该系统置于铅屏蔽箱内,对附近人员的辐射暴露量较低;我们测量到,在[15O]H2O生成的2分钟内,辐射率约为100微希沃特/小时(10毫伦琴/小时),生成3.7吉贝可(100毫居里)后,辐射率小于50微希沃特/小时(小于5毫伦琴/小时)。重复性研究表明,注射活度的变化系数为2%-3%。