[Modern electrodes with and without steroid: effects on stimulation current of cardiac pacemakers].

作者信息

Schuchert A, Kuck K H, Meinertz T

机构信息

Medizinische Universitäts- und Poliklinik, Universitäts-Krankenhaus Eppendorf, Hamburg.

出版信息

Z Kardiol. 1995 Apr;84(4):289-95.

PMID:7785301
Abstract

The aim of the investigation was to test whether new leads without steroid, a meshwire tip leads (Ionyx 4180, CPI; n = 10) and a carbon tip lead (Facet ITP 13, Vitatron; n = 10), have electrical characteristics similar to a new steroid-eluting tip lead (CapSure SP 4023, Medtronic; n = 10) and their impact on the pacemaker's pacing current. Pacing thresholds, impedance, and R-wave amplitudes measured at implantation were similar for the three leads. One, 4 and 12 weeks after implantation both nonsteroid leads had significantly higher pacing thresholds at 0.1 and 0.3 ms pulse duration in comparison to the steroid lead (after 12 weeks at 0.3 ms steroid: 0.9 +/- 0.2 V, carbon: 2.1 +/- 0.5 V; meshwire: 1.5 +/- 0.5 V). This result was restricted to the carbon lead after 52 weeks. At 0.5 ms pulse duration higher pacing thresholds were obtained for the carbon lead (after 12 weeks at 0.5 ms; steroid: 0.8 V, carbon: 1.7 +/- 0.6 V; meshwire: 1.1 +/- 0.4 V). Impedance of the steroid lead was 531 +/- 61 ohms, 535 +/- 54 ohms, and 511 +/- 50 ohms, respectively, after 4, 12, and 52 weeks, whereas the carbon lead had significantly higher values with 652 +/- 84 ohms, 669 +/- 93 ohms, and 657 +/- 107 ohms, respectively. The impedance of the meshwired lead (4 weeks: 585 +/- 92 ohms; 12 weeks: 592 +/- 89 ohms; 52 weeks: 550 +/- 130 ohms) did not differ from the steroid lead.(ABSTRACT TRUNCATED AT 250 WORDS)

摘要

文献AI研究员

20分钟写一篇综述,助力文献阅读效率提升50倍。

立即体验

用中文搜PubMed

大模型驱动的PubMed中文搜索引擎

马上搜索

文档翻译

学术文献翻译模型,支持多种主流文档格式。

立即体验