对比度均衡对给予患者的能量的影响:两种牙科X射线发生器与两种口腔内胶片类型的比较

Effects of contrast equalization on energy imparted to the patient: a comparison of two dental generators and two types of intraoral film.

作者信息

Helmrot E, Carlsson G A, Eckerdal O

机构信息

Department of Radiation Physics, Faculty of Health Sciences, The Medical School, Linköping, Sweden.

出版信息

Dentomaxillofac Radiol. 1994 May;23(2):83-90. doi: 10.1259/dmfr.23.2.7835508.

Abstract

Technical evolution in maxillofacial radiology has in the last decade provided faster films and the constant potential generator. The consequences of these innovations for radiographic contrast and energy imparted to the patient are analysed. On the basis of physical measurements a test model has been developed for correcting exposure parameters in order to maintain or restore image contrast. These measurements are expressed in and developed from basic radiological concepts and physical formulas presented in an earlier paper (Helmrot E. et al., Dentomaxillofac. Radiol. 1991; 20: 135-46). The test model can also be used to demonstrate the balance between contrast and energy imparted to the patient in the radiographic process. Changing to constant potential generators and faster film may each result in a degradation in contrast, which is possible to restore by a controlled adjustment of the kV-setting. Maintenance of constant image quality results in a slight reduction in the net gain in energy imparted, due to the generator and/or film shift. When, for example, a conventional single-pulse generator operated at 65kVp tube potential was replaced by a modern constant potential unit, the kV-setting had to be decreased by 5 to 8 kV to maintain the same radiographic contrast. This correction could be done without increasing energy imparted to the patient, taking into account the fact that the spectral characters of the photon energy are not identical. If, in addition, faster intraoral film with lower film contrast was introduced, together with the constant potential unit, the kV-setting had to be further decreased to maintain the radiographic contrast.(ABSTRACT TRUNCATED AT 250 WORDS)

摘要

在过去十年中,颌面放射学的技术发展带来了成像更快的胶片和恒压发生器。本文分析了这些创新对射线照相对比度以及给予患者的能量的影响。基于物理测量,开发了一种测试模型,用于校正曝光参数,以维持或恢复图像对比度。这些测量是根据早期一篇论文(Helmrot E.等人,《牙颌面放射学》,1991年;20:135 - 46)中提出的基本放射学概念和物理公式得出并表示的。该测试模型还可用于展示射线照相过程中对比度与给予患者的能量之间的平衡。更换为恒压发生器和更快的胶片可能各自导致对比度下降,通过对千伏设置进行可控调整有可能恢复。由于发生器和/或胶片的改变,维持恒定的图像质量会导致给予的能量净增益略有降低。例如,当将在65kVp管电压下运行的传统单脉冲发生器替换为现代恒压装置时,千伏设置必须降低5至8kV才能维持相同的射线照相对比度。考虑到光子能量的光谱特性不同,在不增加给予患者能量的情况下可以进行这种校正。此外,如果与恒压装置一起引入具有较低胶片对比度的更快的口内胶片,则必须进一步降低千伏设置以维持射线照相对比度。(摘要截取自250字)

文献AI研究员

20分钟写一篇综述,助力文献阅读效率提升50倍。

立即体验

用中文搜PubMed

大模型驱动的PubMed中文搜索引擎

马上搜索

文档翻译

学术文献翻译模型,支持多种主流文档格式。

立即体验