用于人类前列腺中ProLase II侧向发射输送系统的钕:钇铝石榴石激光剂量测定法

Neodymium: yttrium-aluminum-garnet laser dosimetry for the prolase II side-firing delivery system in the human prostate.

作者信息

Kabalin J N, Sellers R, Bite G

机构信息

Urology Section of the Veterans Affairs Medical Center, Palo Alto, California.

出版信息

Urology. 1995 Feb;45(2):248-52. doi: 10.1016/0090-4295(95)80013-1.

Abstract

OBJECTIVES

The objective of this study was to define optimal treatment parameters and quantitative dosimetry for neodymium:yttrium-aluminum-garnet (Nd:YAG) laser coagulation prostatectomy utilizing the ProLase II side-firing laser delivery system.

METHODS

Transurethral Nd:YAG laser application was performed with the ProLase II side-firing laser fiber in adult human prostates prior to planned radical prostatectomy. Depth and volume of prostatic tissue coagulation for single, continuous laser applications were measured at variable power settings from 40 to 80 W, while holding total energy delivery constant. Holding the power setting constant, the extent of tissue coagulation was similarly measured for variable treatment times from 45 to 90 seconds.

RESULTS

Peak tissue coagulation for the ProLase II was observed at 60 W--up to a maximum of 13 mm tissue penetration and 3.92 cc volume coagulation following a single spot laser application. The mean depth of tissue coagulation at 60 W power setting was 12.7 mm, with a mean volume of tissue coagulation of 3.07 cc. Holding the power setting constant at 60 W, and increasing treatment time from 45 to 60 seconds, the extent to tissue coagulation increased significantly. However, beyond 60 seconds continuous laser application at 60 W, a plateau in tissue effects was observed, with minimal increase in tissue coagulation between 60 and 90 seconds.

CONCLUSIONS

Using the ProLase II side-firing laser fiber, 60-second Nd:YAG spot laser applications at 60 W power setting represent optimal treatment parameters. The quantitative dosimetry curves produced for the ProLase II fiber are compared with prior dosimetry studies with the Urolase delivery system.

摘要

目的

本研究的目的是确定使用ProLase II侧向发射激光传输系统进行钕:钇铝石榴石(Nd:YAG)激光凝固前列腺切除术的最佳治疗参数和定量剂量测定法。

方法

在计划进行根治性前列腺切除术之前,使用ProLase II侧向发射激光光纤对成年男性前列腺进行经尿道Nd:YAG激光治疗。在40至80 W的可变功率设置下测量单次连续激光应用时前列腺组织凝固的深度和体积,同时保持总能量输送恒定。在功率设置恒定的情况下,类似地测量45至90秒的可变治疗时间的组织凝固程度。

结果

ProLase II的峰值组织凝固在60 W时观察到——单次点激光应用后组织穿透最大可达13 mm,体积凝固为3.92 cc。在60 W功率设置下组织凝固的平均深度为12.7 mm,组织凝固的平均体积为3.07 cc。将功率设置恒定在60 W,并将治疗时间从45秒增加到60秒,组织凝固程度显著增加。然而,在60 W连续激光应用超过60秒后,观察到组织效应达到平稳状态,在60至90秒之间组织凝固增加最小。

结论

使用ProLase II侧向发射激光光纤,在60 W功率设置下进行60秒的Nd:YAG点激光应用代表最佳治疗参数。将ProLase II光纤产生的定量剂量测定曲线与先前使用Urolase传输系统的剂量测定研究进行比较。