[Measurement of bone density in the wrist using X-ray absorptiometry: comparison with measurements of other sites].

作者信息

Rey P, Sornay-Rendu E, Garnero P, Vey-Marty B, Delmas P D

机构信息

Unité INSERM 403, Hôpital Edouard Herriot, France.

出版信息

Rev Rhum Ed Fr. 1994 Oct;61(9):619-26.

PMID:7858595
Abstract

In order to validate a new technique to measure bone mineral density (BMD) of the forearm with a single X-ray absorptiometry device (DTX 100), we have measured the BMD of 195 women, aged 32 to 83 years, free of metabolic bone disease, simultaneously on two devices: 1. BMD of the ultradistal radius and of the distal radius and ulna on DTX 100, Osteometer, 2. BMD of the lumbar spine, of the proximal femur, of the whole body and of the distal and ultradistal radius by dual X-ray absorptiometry (DXA) on an Hologic 2000 device. The precision of DTX, assessed by 5 repeated measurements on 9 volunteers, was 1% for the distal site and 2% for the ultradistal site. There was a high correlation (r = 0.93) between radial BMD measurement by DTX and Hologic 2000. Correlations between forearm BMDs measured by DTX and whole body BMD were as high (r = 0.80-0.82) as between whole body BMD and lumbar spine (r = 0.82), femoral neck (r = 0.77) and femoral trochanter (r = 0.76). Age-related bone loss measured by DTX was 27% and 31% at the distal and ultra distal radius respectively, as compared to 17% for the lumbar spine, 21% for the femoral neck and 16% for the whole body BMD. In a group of 20 untreated osteoporotic women with vertebral fractures, forearm BMD was as sensitive (T score of -3.1 and -2.9 for distal and ultra distal) as was lumbar spine (T = -2.9) and femoral neck (T = -2.5).(ABSTRACT TRUNCATED AT 250 WORDS)

摘要

为验证使用单能X线吸收测定仪(DTX 100)测量前臂骨矿物质密度(BMD)的新技术,我们对195名年龄在32至83岁、无代谢性骨病的女性,同时使用两种设备测量其BMD:1. 使用DTX 100骨密度仪测量尺骨远极及桡骨远端和尺骨的BMD;2. 使用Hologic 2000设备通过双能X线吸收法(DXA)测量腰椎、股骨近端、全身以及桡骨远端和远极的BMD。通过对9名志愿者进行5次重复测量评估DTX的精密度,远端部位为1%,远极部位为2%。DTX测量的桡骨BMD与Hologic 2000测量结果高度相关(r = 0.93)。DTX测量的前臂BMD与全身BMD之间的相关性(r = 0.80 - 0.82)与全身BMD和腰椎(r = 0.82)、股骨颈(r = 0.77)以及股骨转子(r = 0.76)之间的相关性一样高。与腰椎的17%、股骨颈的21%以及全身BMD的16%相比,DTX测量的桡骨远端和远极与年龄相关的骨质流失分别为27%和31%。在一组20名未经治疗的有椎体骨折的骨质疏松女性中,前臂BMD与腰椎(T值为 -2.9)和股骨颈(T值为 -2.5)一样敏感(远端和远极的T值分别为 -3.1和 -2.9)。(摘要截取自250字)

文献检索

告别复杂PubMed语法,用中文像聊天一样搜索,搜遍4000万医学文献。AI智能推荐,让科研检索更轻松。

立即免费搜索

文件翻译

保留排版,准确专业,支持PDF/Word/PPT等文件格式,支持 12+语言互译。

免费翻译文档

深度研究

AI帮你快速写综述,25分钟生成高质量综述,智能提取关键信息,辅助科研写作。

立即免费体验