Accuracy, alarm limits and rise times of 12 oxygen analysers.

作者信息

Lauber R, Steiner A, Zbinden A M

机构信息

Research Section, University Hospital of Berne, Switzerland.

出版信息

Anaesthesia. 1994 Dec;49(12):1071-7. doi: 10.1111/j.1365-2044.1994.tb04359.x.

Abstract

The Comité Européen de Normalisation recently proposed a new standard for 'the particular requirements of oxygen monitors for medical use'. The feasibility of this proposed standard was tested in respect of (1) accuracy of alarm activation (2) accuracy of oxygen display value during both continuous and cyclical gas flows (3) rise time during rapid changes in oxygen concentration in the following 12 analysers: Datex Capnomac II and Servomex 570A (paramagnetic); Brüel & Kjaer 1304 (magnetoacoustic); Criticare Poet II, Multinex, Dräger Oxydig, Dräger PM 8030, Megamed 046A (part of the Megamed 700 ventilator), Ohmeda 5120, Spacelabs Multigas, Teledyne TED 200 (galvanic); Kontron OM 810 (polarographic). All the analysers tested displayed an oxygen reading which was within +/- 3 vol% of the actual oxygen concentrations of the test gases (15, 21, 40, 60 and 100 vol%). A cyclical pressure of between -1.5 to +8 kPa did not affect the measured oxygen concentration as displayed by the Brüel & Kjaer 1304, Datex Capnomac II and Servomex 570A analysers. The remainder, however, showed, depending on their measuring principle, a display error of between -1 and +6 vol%. After exposure to high pressure all the oximeters functioned normally. Some of the tested devices showed more than 2% of deviation between their alarm activation and the preset alarm limits. Only the Kontron OM 810, the Megamed 046A and the Spacelabs Multigas monitors satisfied the requirements at all the tested oxygen concentrations. The time required by the oxygen analyser to display the rise from 29 to 92 vol % after a sudden change of concentration from 21 to 100 vol % O2 is defined as "rise time" and must not, according to the Comité Européen de Normalisation standard proposal, exceed the manufacturers' specification by more than a factor of 1.15.(ABSTRACT TRUNCATED AT 250 WORDS)

摘要

文献检索

告别复杂PubMed语法,用中文像聊天一样搜索,搜遍4000万医学文献。AI智能推荐,让科研检索更轻松。

立即免费搜索

文件翻译

保留排版,准确专业,支持PDF/Word/PPT等文件格式,支持 12+语言互译。

免费翻译文档

深度研究

AI帮你快速写综述,25分钟生成高质量综述,智能提取关键信息,辅助科研写作。

立即免费体验