Precision and accuracy of measurements of whole-body bone mineral: comparisons between Hologic, Lunar and Norland dual-energy X-ray absorptiometers.

作者信息

Tothill P, Avenell A, Reid D M

机构信息

Department of Medical Physics and Medical Engineering, University of Edinburgh, Western General Hospital, UK.

出版信息

Br J Radiol. 1994 Dec;67(804):1210-7. doi: 10.1259/0007-1285-67-804-1210.

Abstract

Measurements of whole-body bone mineral made by Hologic, Lunar and Norland dual-energy X-ray absorptiometers have been compared. It was found that in each case the results were changed by new software protocols introduced by the manufacturers during the course of the study. With a moderately anthropomorphic model, the later software corrected some anomalies of regional bone mineral content (BMC) observed earlier. There was some slight dependence of total BMC on thickness and fat proportion and up to 15% difference between instruments. Measurements on volunteers showed good precision, but there were differences between instruments made by different manufacturers. There were high correlations, but the slopes of regression lines suggested differences of calibration of up to 8%; the standard errors of the estimates were 110 to 190 g. with maximum deviations from regression of 17%. There were regional disparities in BMC, particularly in the trunk, which arise (in part at least) from the imposition of a higher bone threshold by Hologic. From the pattern of results it was concluded that different assumptions were made by the manufacturers, particularly concerning the fat distribution model, which preclude the interchangeability of results from different instruments.

摘要

文献检索

告别复杂PubMed语法,用中文像聊天一样搜索,搜遍4000万医学文献。AI智能推荐,让科研检索更轻松。

立即免费搜索

文件翻译

保留排版,准确专业,支持PDF/Word/PPT等文件格式,支持 12+语言互译。

免费翻译文档

深度研究

AI帮你快速写综述,25分钟生成高质量综述,智能提取关键信息,辅助科研写作。

立即免费体验