An evaluation of the assess peak flow meter on human volunteers.

作者信息

Venkatesan E A, Walter S, Ray D

机构信息

Department of Physiology, Christian Medical College, Vellore.

出版信息

Indian J Physiol Pharmacol. 1994 Oct;38(4):285-8.

PMID:7883294
Abstract

The Assess peak flow meter was tested by comparing PEF values obtained on it for 49 normal subjects, with the values obtained for the same subjects on the Wright and mini-Wright meters. Its reproducibility was tested by comparing its coefficient of variation with that of the other 2 instruments. PEF values on the Assess were 5% lower in lower ranges and 9% higher in higher ranges tested as compared with those on the Wright. Values from the mini-Wright and Wright paralleled each other with the former values 3-5% higher. Variabilities on the Assess were 20-50% higher than those of the other 2 meters. Further, PEF values > 670 L/min could not be read by the Assess. Although the Assess fulfils necessary criteria for accuracy and reproducibility, it appears to be less reliable than the other 2 instruments, and its range limits and utility.

摘要

通过比较49名正常受试者使用Assess峰值流量计测得的呼气峰流速(PEF)值与使用Wright和微型Wright流量计为同一受试者测得的值,对Assess峰值流量计进行了测试。通过将其变异系数与其他两种仪器的变异系数进行比较,测试了其再现性。与Wright流量计相比,在较低范围测试时,Assess上的PEF值低5%,在较高范围测试时高9%。微型Wright流量计和Wright流量计的值相互平行,前者的值高3-5%。Assess流量计的变异性比其他两种流量计高20-50%。此外,Assess流量计无法读取大于670升/分钟的PEF值。尽管Assess流量计符合准确性和再现性的必要标准,但它似乎不如其他两种仪器可靠,且有其范围限制和实用性。