Multicenter assessment of the QuietTrak ambulatory blood pressure recorder according to the 1992 AAMI guidelines.

作者信息

White W B, Susser W, James G D, Marra L, McCabe E J, Pickering T G, Streeten D H

机构信息

Section of Hypertension and Vascular Diseases, University of Connecticut Health Center, Farmington 06032-3940.

出版信息

Am J Hypertens. 1994 Jun;7(6):509-14. doi: 10.1093/ajh/7.6.509.

Abstract

To clinically evaluate the QuietTrak ambulatory blood pressure (BP) recorder (Tycos-Welch-Allyn, Arden, NC), we assessed the device according to the performance criteria set out by the 1992 Association for the Advancement of Medical Instrumentation (AAMI) guidelines. The QuietTrak is a portable, noninvasive recorder that uses an auscultatory measuring system. As recommended in the 1992 AAMI guidelines, a large, heterogeneous population was recruited for the study using three clinic sites to assess accuracy and clinical performance. In addition, observer agreement and the effects of age, arm circumference, heart rate, posture, and blood pressure level on the observer-device differences were analyzed. There were 1098 simultaneous, same arm BP measurements performed in 122 subjects by the QuietTrak recorder versus two skilled clinicians per site using a teaching stethoscope and 24-h blood pressure recordings performed in 46 subjects to assess reliability of the monitor. The differences in observers for all sites were 0.7/-0.5 +/- 3.8/3.5 mm Hg with 89% of the BPs within 5 mm Hg. The mean difference between observer average and the device (all positions) was 0.3/-1.5 +/- 5.0/7.5 mm Hg. The agreement between the QuietTrak and mercury column determinations was within 10 mm Hg for 92 to 94% of systolic readings and 91 to 93% of diastolic readings, depending on the posture. The limits of agreement (2 standard deviations about the mean difference) between observers and the device for systolic BP tended to be lower for the sitting position (-11 to 10 mm Hg) compared to supine (-14 to 13 mm Hg) and standing (-14 to 14 mm Hg) positions.(ABSTRACT TRUNCATED AT 250 WORDS)

摘要

为了对QuietTrak动态血压记录仪(泰科思-韦尔奇-阿林公司,北卡罗来纳州阿登)进行临床评估,我们根据1992年医疗仪器促进协会(AAMI)指南规定的性能标准对该设备进行了评估。QuietTrak是一种便携式、非侵入性记录仪,采用听诊测量系统。按照1992年AAMI指南的建议,我们利用三个临床地点招募了一个大型、异质性人群进行研究,以评估其准确性和临床性能。此外,还分析了观察者之间的一致性以及年龄、臂围、心率、姿势和血压水平对观察者与设备差异的影响。QuietTrak记录仪在122名受试者中进行了1098次同侧手臂同步血压测量,每个地点有两名技术熟练的临床医生使用教学听诊器进行测量,并在46名受试者中进行了24小时血压记录,以评估该监测仪的可靠性。所有地点观察者之间的差异为0.7/-0.5 +/- 3.8/3.5毫米汞柱,89%的血压值在5毫米汞柱以内。观察者平均值与设备(所有体位)之间的平均差异为0.3/-1.5 +/- 5.0/7.5毫米汞柱。根据姿势不同,QuietTrak与汞柱测定值之间的一致性在收缩压读数的92%至94%以及舒张压读数的91%至93%范围内,相差在10毫米汞柱以内。观察者与设备之间收缩压的一致性界限(平均差异的2个标准差)在坐位时(-11至10毫米汞柱)往往低于仰卧位(-14至13毫米汞柱)和站立位(-14至14毫米汞柱)。(摘要截取自250字)

文献AI研究员

20分钟写一篇综述,助力文献阅读效率提升50倍。

立即体验

用中文搜PubMed

大模型驱动的PubMed中文搜索引擎

马上搜索

文档翻译

学术文献翻译模型,支持多种主流文档格式。

立即体验