Schmidt J, Nordmann K
Klinik und Poliklinik für Orthopädie, Universität zu Köln, Germany.
Med Biol Eng Comput. 1994 May;32(3):258-60. doi: 10.1007/BF02512520.
The removal of bone cement from the femoral canal in the case of hip prosthesis loosening takes a considerable amount of time and carries the additional risk of complications, thus risking bone damage and perhaps even endangering the successful reimplantation of a new prosthesis. We have demonstrated experimentally that the milling of bone and bone cement produces acoustic emission waves of different intensity. These waves can be measured by a commercial acoustic emission transducer. This method enables us to control a milling device during the removal of bone cement from deep in the femoral canal, avoiding the serious damage to the bone and thus enabling a new prosthesis to be reimplanted.