Oh I, Bourne R B, Harris W H
J Bone Joint Surg Am. 1983 Dec;65(9):1335-8.
Aseptic loosening is the most common cause of femoral component failure in total hip arthroplasty. Even with new cementing techniques, bone-cement pressurization is low in the most proximal area of the femoral canal. To correct this problem, we devised the femoral cement compactor. This device increased the pressurization of bone cement in the proximal area of the femur by 71 per cent and increased the duration of the pressurization fourfold as compared with the results achieved by finger-packing methods. Pressurization of the cement was also increased in the middle and distal areas of the proximal portion of the femoral canal. These findings suggest that using the femoral cement compactor may contribute to improved cement fixation of the femoral component during total hip replacement.
无菌性松动是全髋关节置换术中股骨部件失效的最常见原因。即使采用新的骨水泥固定技术,在股骨髓腔最近端区域的骨水泥加压也很低。为纠正这一问题,我们设计了股骨骨水泥压实器。与手指填塞法相比,该装置使股骨近端区域的骨水泥压力增加了71%,并使加压持续时间增加了四倍。在股骨髓腔近端的中部和远端区域,骨水泥的压力也有所增加。这些发现表明,在全髋关节置换术中使用股骨骨水泥压实器可能有助于改善股骨部件的骨水泥固定。