Cook L T, Insana M F, McFadden M A, Hall T J, Cox G G
University of Kansas Medical Center, Department of Diagnostic Radiology, Kansas City 66160-7234.
Med Phys. 1994 May;21(5):691-5. doi: 10.1118/1.597172.
Contrast-detail (CD) analysis was used to compare the low-contrast detectability of computed radiography (CR) and screen-film (SF) as applied to the task of adult chest radiography. A phantom was constructed and imaged using the same exposure factors throughout all experiments. Within-observer variance, between-observer variance, and image sample variance were calculated and used to estimate the standard error for each experiment. The results of these CD experiments agreed with the predictions of the Rose model. Observers performed equally well for low-contrast target detection using CR and SF.
使用对比细节(CD)分析来比较计算机X线摄影(CR)和屏-片摄影(SF)在成人胸部X线摄影任务中的低对比度可探测性。构建了一个模体,并在所有实验中使用相同的曝光参数进行成像。计算了观察者内方差、观察者间方差和图像样本方差,并用于估计每个实验的标准误差。这些CD实验的结果与罗斯模型的预测一致。观察者使用CR和SF进行低对比度目标检测的表现相当。