固定不佳对计算机辅助角膜地形图分析的影响。假性圆锥角膜。

The effect of poor fixation on computer-assisted topographic corneal analysis. Pseudokeratoconus.

作者信息

Hubbe R E, Foulks G N

机构信息

Duke University Eye Center, Durham, North Carolina 27710.

出版信息

Ophthalmology. 1994 Oct;101(10):1745-8. doi: 10.1016/s0161-6420(94)31109-2.

Abstract

INTRODUCTION

The effect of changing the alignment of the corneal surface in relation to the plane of the videokeratoscope mires in computer-assisted topographic corneal analysis was evaluated.

METHODS

The corneas of five subjects were analyzed with the visual axis at 0 degrees, 2.5 degrees, 5 degrees, and 10 degrees below the central axis of the videokeratoscope. The relative steepness of the inferior and superior cornea was compared using inferior-superior (I-S) values at each deviation from the central axis. A similar analysis was performed on three aspheric test surfaces with different central curvatures.

RESULTS

The difference in the relative steepness of the inferior and superior cornea was found to increase in proportion to the deviation from alignment with the central axis in both the test subjects and the aspheric test surfaces.

CONCLUSION

This change in relative steepness produces a pattern that mimics keratoconus and can be seen at deviations of less than 5 degrees in some patients. The increase in the I-S value was significant at 5 degrees of deviation (P < 0.05). It also was found that this effect increased as the central curvature of the aspheric surfaces increased.

摘要

引言

评估了在计算机辅助角膜地形图分析中,角膜表面相对于角膜视频镜环曲面平面的对准变化所产生的影响。

方法

对五名受试者的角膜进行分析,视轴分别位于角膜视频镜中心轴下方0度、2.5度、5度和10度处。在每次偏离中心轴时,使用上下(I-S)值比较下角膜和上角膜的相对陡峭程度。对具有不同中心曲率的三个非球面测试表面进行了类似分析。

结果

发现在测试对象和非球面测试表面中,下角膜和上角膜的相对陡峭程度差异均与偏离中心轴对准的程度成比例增加。

结论

这种相对陡峭程度的变化产生了一种类似圆锥角膜的模式,在一些患者中,偏离小于5度时即可观察到。在5度偏离时,I-S值的增加具有显著性(P < 0.05)。还发现,随着非球面表面中心曲率的增加,这种效应也会增强。