Pár A, Mihály I, Kömives K
Pécsi Orvostudományi Egyetem I. Belgyógyászati Klinika.
Orv Hetil. 1994 Sep 25;135(39):2143-6.
An one-step enzyme-linked immunoabsorbent method, named as HBsAg Uniform II has been described for the detection of serum HBsAg, and a comparison was made with a widely used ELISA technique HBsAg Hepanostika test, to evaluate sensitivity, specificity and reproducibility of the method. A total of 531 serum samples from patients with liver disease and with renal failure, as well as 1065 samples from blood donors have been investigated. While the sensitivity of Uniform II vs. Hepanostika was 99.5% vs. 72.7%, the specificity was 99.2% of both methods. The positive predictive values did not differ (99.5% vs. 99.2%), however, the negative predictive values were 99.2% vs. 71.7%, respectively, in favour of Uniform II test. The Uniform II confirmatory test confirmed the positive HBsAg results in 94%, this rate was 74% using Hepanostika system. The new method proved to be a simple, quick, reliable test, which can be useful as a valuable tool in both the clinical diagnosis and blood donor screening.
已描述了一种名为HBsAg Uniform II的一步法酶联免疫吸附法用于检测血清HBsAg,并与广泛使用的ELISA技术HBsAg Hepanostika试验进行了比较,以评估该方法的敏感性、特异性和可重复性。共调查了531份来自肝病和肾衰竭患者的血清样本以及1065份来自献血者的样本。虽然Uniform II相对于Hepanostika的敏感性为99.5%对72.7%,但两种方法的特异性均为99.2%。阳性预测值没有差异(99.5%对99.2%),然而,阴性预测值分别为99.2%对71.7%,支持Uniform II试验。Uniform II确证试验确认了94%的HBsAg阳性结果,使用Hepanostika系统时该比率为74%。新方法被证明是一种简单、快速、可靠的检测方法,可作为临床诊断和献血者筛查的有价值工具。