Tagawa H, Kado M, Okada A, Furukawa H
Department of Ophthalmology, Asashikawa Medical College, Hokkaido, Japan.
Nippon Ganka Gakkai Zasshi. 1994 Aug;98(8):772-6.
Reproducibility of results and possible clinical use of a newly developed noncontact autofocus specular microscope were studied. Morphological parameters of corneal endothelium were analyzed on images taken with the noncontact autofocus specular microscope and the ordinary contact specular microscope simultaneously in 10 eyes of 10 normal subjects. Intraphotographic and interphotographic study by autoanalysis of 5 images each of 10 eyes taken by the noncontact instrument showed small coefficient variation of mean cell area and cell density (less than 5%, individually). These morphological parameters showed about 7% difference between autoanalysis and manual analysis of the images taken by the noncontact instrument. In comparison with the images taken by the contact instrument, the images taken by the noncontact instrument showed about 7% difference in mean cell area and cell density (p < 0.001, individually). In respect of reproducibility, this newly developed noncontact autofocus specular microscope is clinically useful. The differences of morphological parameters between the new and ordinary instruments might be caused the differences in corneal shapes.
研究了新开发的非接触自动对焦角膜内皮显微镜结果的可重复性及其可能的临床应用。在10名正常受试者的10只眼中,同时使用非接触自动对焦角膜内皮显微镜和普通接触式角膜内皮显微镜拍摄图像,分析角膜内皮的形态学参数。对非接触仪器拍摄的10只眼的5张图像进行自动分析的摄影内和摄影间研究显示,平均细胞面积和细胞密度的变异系数较小(个体均小于5%)。这些形态学参数在非接触仪器拍摄图像的自动分析和手动分析之间显示出约7%的差异。与接触式仪器拍摄的图像相比,非接触仪器拍摄的图像在平均细胞面积和细胞密度上显示出约7%的差异(个体p<0.001)。就可重复性而言,这种新开发的非接触自动对焦角膜内皮显微镜在临床上是有用的。新仪器与普通仪器之间形态学参数的差异可能是由角膜形状的差异引起的。