Suppr超能文献

三种半自动角膜内皮显微镜的比较研究

Comparative study of three semiautomated specular microscopes.

作者信息

Landesz M, Siertsema J V, Van Rij G

机构信息

Department of Ophthalmology, University Hospital Groningen, The Netherlands.

出版信息

J Cataract Refract Surg. 1995 Jul;21(4):409-16. doi: 10.1016/s0886-3350(13)80531-7.

Abstract

We compared two clinical video-assisted specular microscopes (Zeiss, noncontact, and the wide-field Keeler Konan sp 3300, contact) with an autofocus microscope (Konan noncon Robo-ca sp 8000, noncontact) with built-in analyzing software by studying the morphometry of the central corneal endothelium of 12 eyes in 12 patients. The mean coefficients of variation of the cell size analysis of three successive images per cornea for the three methods were 3.9%, 4.0%, and 2.2%, respectively. One-way analysis of variance showed a significant difference in the mean cell area measured with the three methods; there was no significant difference in the results of hexagonality (pleomorphism) and coefficient of variation (polymegathism). Although each microscope showed a high reproducibility, we do not recommend using the three interchangeably. The Konan noncon Robo-ca sp 8000 specular microscope provides a rapid morphometric endothelial analysis. The Keeler Konan sp 3300 is better for more detailed studies and photography of the corneal endothelium.

摘要

我们通过研究12例患者12只眼的中央角膜内皮形态学,将两台临床视频辅助角膜内皮显微镜(蔡司非接触式和宽视野基勒科南sp 3300接触式)与一台带有内置分析软件的自动对焦显微镜(科南非接触式Robo-ca sp 8000非接触式)进行了比较。三种方法对每个角膜连续三张图像的细胞大小分析的平均变异系数分别为3.9%、4.0%和2.2%。单因素方差分析显示,三种方法测量的平均细胞面积存在显著差异;六角形(多形性)和变异系数(大小不均一性)的结果无显著差异。尽管每种显微镜都具有很高的重复性,但我们不建议将这三种显微镜互换使用。科南非接触式Robo-ca sp 8000角膜内皮显微镜可提供快速的内皮形态学分析。基勒科南sp 3300更适合对角膜内皮进行更详细的研究和摄影。

文献AI研究员

20分钟写一篇综述,助力文献阅读效率提升50倍。

立即体验

用中文搜PubMed

大模型驱动的PubMed中文搜索引擎

马上搜索

文档翻译

学术文献翻译模型,支持多种主流文档格式。

立即体验