Schmid K M, Voelker W, Mewald J, Paul H J, Wehrmann M, Bültmann B, Karsch K R
Department of Cardiology, University of Tübingen, FRG.
Basic Res Cardiol. 1994;89 Suppl 1:129-35. doi: 10.1007/978-3-642-85660-0_12.
To evaluate the impact and limitations of intracoronary ultrasound in the assessment of lumen, we examined 80 segments of 20 isolated coronary arteries with a mechanical ultrasound device (CVIS) comparing the results of ultrasound with the corresponding histological specimens. Ultrasound was performed with and without echogenic contrast dye (Laevovist, Schering AG, FRG). After application of contrast dye, correlation of luminal area between histology and ultrasound was improved from r = 0.85 to r = 0.89 (p = ns). Accuracy of lumen measurements was low in segments < 2.5 mm; only after application of contrast dye a relationship between ultrasound and histological measurements was found. In all cases in which a deviation of more than 20% between ultrasound and histology was observed, this deviation could be reduced by the application of contrast dye. There are considerable limitations in the accuracy of ultrasound measurements in the near field. Thus, further improvement of intracoronary ultrasound devices is mandatory. However, with the use of the currently available systems, additional application of echogenic contrast dye can improve accuracy of luminal measurements, especially in smaller size vessels.
为评估冠状动脉内超声在评估管腔方面的影响和局限性,我们使用机械超声设备(CVIS)检查了20条离体冠状动脉的80个节段,将超声检查结果与相应的组织学标本进行比较。超声检查在使用和不使用回声造影剂(Levovist,先灵公司,德国)的情况下进行。应用造影剂后,组织学与超声检查之间管腔面积的相关性从r = 0.85提高到r = 0.89(p =无显著性差异)。管腔小于2.5 mm节段的管腔测量准确性较低;仅在应用造影剂后才发现超声与组织学测量之间存在相关性。在所有观察到超声与组织学之间偏差超过20%的病例中,应用造影剂可减少这种偏差。近场超声测量的准确性存在相当大的局限性。因此,必须进一步改进冠状动脉内超声设备。然而,使用目前可用的系统,额外应用回声造影剂可提高管腔测量的准确性,特别是在较小尺寸的血管中。