Schmid K M, Mewald J, Voelker W, Paul H J, Wehrmann M, Bültmann B, Karsch K R
Department of Cardiology, University of Tuebingen, Germany.
Am J Cardiol. 1993 Jan 15;71(2):184-91. doi: 10.1016/0002-9149(93)90736-v.
To evaluate the impact and limitations of intracoronary ultrasound, 64 segments of 13 isolated coronary arteries were examined and 2 mechanical devices (device A, 30 MHz 5Fr [CVIS] and device B, 20 MHz 4.8Fr [Diasonics/Boston Scientific] were compared with the corresponding histologic specimens. Luminal dimensions were assessed with and without echogenic contrast dye (Laevovist, Schering AG, Germany). After application of contrast dye, correlation of luminal area between histology and ultrasound was improved from r = 0.62 to r = 0.77 (device B; p < 0.05) and from r = 0.82 to r = 0.88 (device A; p = NS). Low accuracy of lumen measurements in segments < 2.5 mm could be improved by application of contrast dye. The number of quadrants in which wall thickness measurements were impossible was significantly higher for device B (n = 56; 22%) than for device A (n = 28; 11%); p < 0.01. This may be due to the different ringdown diameters of both systems (B, 2.6 mm; A, 2.0 mm; p < 0.0001). In assessing wall thickness only in segments of > 2.5 mm, a reliable correlation between ultrasound and histology was found (A, r = 0.80; B, r = 0.60). Sensitivity of plaque (n = 51) detection was lower for device B (63%) than for device A (82%, p < 0.05), and measurements correlated with histology only for device A. There are considerable differences in the accuracy of ultrasound measurements between mechanical systems. Nevertheless, additional application of contrast dye can improve accuracy of luminal measurements, especially in smaller vessels.
为评估冠状动脉内超声的影响和局限性,对13条离体冠状动脉的64个节段进行了检查,并将2种机械装置(装置A,30MHz 5Fr[CVIS]和装置B,20MHz 4.8Fr[Diasonics/波士顿科学公司])与相应的组织学标本进行了比较。使用和不使用回声造影剂(Levovist,先灵葆雅公司,德国)评估管腔尺寸。应用造影剂后,装置B的组织学与超声之间管腔面积的相关性从r = 0.62提高到r = 0.77(p < 0.05),装置A从r = 0.82提高到r = 0.88(p =无显著性差异)。应用造影剂可提高直径<2.5mm节段管腔测量的准确性。装置B无法进行壁厚测量的象限数量(n = 56;22%)显著高于装置A(n = 28;11%);p < 0.01。这可能是由于两个系统的振铃直径不同(B,2.6mm;A,2.0mm;p < 0.0001)。仅在直径>2.5mm的节段评估壁厚时,发现超声与组织学之间存在可靠的相关性(A,r = 0.80;B,r = 0.60)。装置B对斑块(n = 51)的检测灵敏度(63%)低于装置A(82%,p < 0.05),且仅装置A的测量结果与组织学相关。机械系统之间超声测量的准确性存在相当大的差异。然而,额外应用造影剂可提高管腔测量的准确性,尤其是在较小的血管中。