用于磁共振引导下针定位的原型乳腺线圈。

Prototype breast coil for MR-guided needle localization.

作者信息

Heywang-Köbrunner S H, Huynh A T, Viehweg P, Hanke W, Requardt H, Paprosch I

机构信息

Department of Diagnostic Radiology, Martin Luther University of Halle-Wittenberg, Germany.

出版信息

J Comput Assist Tomogr. 1994 Nov-Dec;18(6):876-81. doi: 10.1097/00004728-199411000-00006.

Abstract

OBJECTIVE

Evaluation of breast lesions detected by MR alone is a problem: Preoperative MR-guided localization is difficult, since the breast is not accessible by a needle within the breast coil. Magnetic resonance-guided core biopsy and needle aspiration are impossible. Together with Siemens Erlangen, a prototype localization breast coil has been developed and tested.

MATERIALS AND METHODS

The device consists of a circular polarized coil, which contains two plates for mediolateral breast compression. They are perforated with numerous holes and thus allow access to the breast from both sides. By means of bushings, which fit into the holes, sterile needle insertion in a horizontal path is possible.

RESULTS

So far, precise needle insertion has been possible in 10 of 11 lesions, allowing exact needle insertion into 5 invasive carcinomas (1 MR-detected), 1 MR-detected additional focus, 2 in situ carcinomas (1 MR-detected), and 3 benign lesions.

CONCLUSION

Our studies show that MR-guided needle localization is possible.

摘要

目的

单独通过磁共振成像(MR)检测乳腺病变存在一个问题:术前MR引导下的定位很困难,因为在乳腺线圈内无法用针进入乳腺。磁共振引导下的粗针活检和针吸活检是不可能的。我们与西门子埃尔朗根公司共同开发并测试了一种原型定位乳腺线圈。

材料与方法

该装置由一个圆极化线圈组成,其中包含两块用于乳腺内外侧压迫的板。它们上有许多孔,因此可以从两侧进入乳腺。通过插入孔中的套管,可以在水平路径上进行无菌针穿刺。

结果

到目前为止,在11个病变中的10个中已实现精确针穿刺,能够将针准确插入5例浸润性癌(其中1例为MR检测到的)、1个MR检测到的额外病灶、2例原位癌(其中1例为MR检测到的)以及3个良性病变中。

结论

我们的研究表明MR引导下的针定位是可行的。