Kronenberg F, Lobentanz E M, König P, Utermann G, Dieplinger H
Institute of Medical Biology and Human Genetics, Innsbruck, Austria.
J Lipid Res. 1994 Jul;35(7):1318-28.
This study investigated the influence of long-term storage, for periods up to 24 months, and multiple freezing and thawing on the measured values of lipoprotein[a] (Lp[a]), apolipoproteins B and A-IV, total and high density lipoprotein (HDL) cholesterol and triglycerides using plasma samples stored at -80 degrees C, -20 degrees C, and 4 degrees C. Samples stored at -80 degrees C or -20 degrees C showed significant changes in Lp[a] after 24 months, with a mean decrease of 7% and 13%, respectively (P < 0.01). The major part of the decrease occurred during the first freezing and thawing. In contrast, apolipoproteins B and A-IV decreased continuously over time (P < 0.05). The increase in plasma concentrations of total and HDL cholesterol and triglycerides was small but significant because of its uniformity. Multiple freezing and thawing influenced only the measured values of Lp[a] and apolipoprotein B. Comparison of samples stored at -80 degrees C and -20 degrees C showed no difference in any of the parameters at any time with the exception of Lp[a] after 18 and 24 months (P < 0.05). After a storage period of 24 months, immunoblotting with detection of apo[a] was possible from samples under each storage condition. ApoB and apoA-IV were detectable only in samples stored at -20 degrees C or -80 degrees C. These data, when compared to recent studies, suggest a critical role of the assay methodology in the reproducibility of measured Lp[a] and apolipoprotein plasma concentrations. We therefore recommend the examination of each system for measurement of long-term stored plasma samples.
本研究调查了长达24个月的长期储存以及多次冻融对脂蛋白[a](Lp[a])、载脂蛋白B和A-IV、总胆固醇和高密度脂蛋白(HDL)胆固醇以及甘油三酯测量值的影响,使用储存在-80℃、-20℃和4℃的血浆样本。储存在-80℃或-20℃的样本在24个月后Lp[a]出现显著变化,平均分别下降7%和13%(P<0.01)。下降的主要部分发生在首次冻融期间。相比之下,载脂蛋白B和A-IV随时间持续下降(P<0.05)。由于其一致性,总胆固醇、HDL胆固醇和甘油三酯的血浆浓度升高幅度较小但具有显著性。多次冻融仅影响Lp[a]和载脂蛋白B的测量值。储存在-80℃和-20℃的样本比较显示,除18个月和24个月后的Lp[a]外,任何时候任何参数均无差异(P<0.05)。储存24个月后,在每种储存条件下的样本均可通过免疫印迹法检测载脂蛋白[a]。载脂蛋白B和载脂蛋白A-IV仅在储存在-20℃或-80℃的样本中可检测到。与近期研究相比,这些数据表明检测方法在测量Lp[a]和载脂蛋白血浆浓度的可重复性中起关键作用。因此,我们建议对每个用于测量长期储存血浆样本的系统进行检测。